MOS OPT100 Instruction Manual Download Page 1

INSTRUCTION MANUAL

OPT100

Oxygen Monitor

Summary of Contents for OPT100

Page 1: ...INSTRUCTION MANUAL OPT100 Oxygen Monitor...

Page 2: ...ressure Compensation 811 mPA 1216mPA Signal Output Analog 4 20 mA programmable Galvanic Isolation 2000 VAC optokupler Line Resistance 1K2 Ohms 2 Sensor lifetime vary with the operating conditions It i...

Page 3: ...g installation Location should be selected carefully to prevent any direct air flow to the Monitor from other machinery and or ventilation Do not install the unit nearby any heat source 1 Electrical c...

Page 4: ...wer should be turned on after finishing the installation completely Remove the electric connection part cover Pay caution to the data cable in between cover and the box and remove this cable from its...

Page 5: ...ET SET 209 o2 SET Adjust HIGH Alarm value with up and down buttons You can set it in between 5 0 to 30 00 aLrL 250 Low O2 alarm value can be set with ALRL parameter Press button Press button Press but...

Page 6: ...ow mentioned steps 1 Connect calibration fitting into N2 bottle 99 9 N2 with pneumatic hose 2 Connect calibration fitting into the O2 sensor via installing into the sensor inlet 3 Open the N2 bottle v...

Page 7: ...ously Be sure that Calibration fitting is connected to the sensor and N2 bottle valve is opened with 99 9 N2 gas supply to the sensor After reaching to 0 value press SET button and stop the gas supply...

Page 8: ...active with this way and start to measure correctly after installation into the OPT100 panel Do not forget to note sensor installation date onto the panel Please check SENSOR CALIBRATION part for new...

Page 9: ...nsor is damaged or expired If value stay constant consult supplier Panel screen isn t working Please check the cable and its socket in between front panel and mainboard and be sure its properly connec...