DEVICE SPECIFICATIONS
52
Power Requirements & Battery
Disposable Battery Type
Three (3) AA alkaline batteries
Disposable Battery Life
12 Hours
Rechargeable Battery Type
Rechargeable Lithium-Ion battery pack
Rechargeable Battery Run-Time
20 Hours
Battery Charging Time
8 Hours
ECG Specifications
ECG
12-lead ECG with specific Signal Acquisition Module (SAM10)
7-lead ECG with specific Signal Acquisition Module (SAM 5)
Simultaneous Leads
Available
10-wire: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
5-wire: I, II, III, aVR, aVL, aVF, V
3-wire: I, II, or III
Acquisition Rate:
40,000 samples/s. for pacemaker detection, 500 samples/s. for analysis
Resolution:
.117µV originally, reduced to 2.5 µV for processing and transmission
Dynamic Range
± 350mV
ECG Gain
½x, 1x, 2x
Trace Speed
slow, normal, fast
CMRR:
According to applicable standards (IEC 60601-2-25 and/or IEC 60601-2-27)
Pacemaker Spikes
When used with 4-wire, 5-wire or 10-wire ECG cable
detect and report pacemaker
spikes up to 2 ms duration with 1 samples precision
Max. Auxiliary Patient
Current:
< 10 µA
Frequency Response:
0.05 to 150 Hz
Input Impedance:
>2.5 M Ω at 10 Hz
Electrodes:
Compatible with Ag/AgCl disposable electrodes
HR Report from the Central
Station
Less than one second. Heart rate is only available when the S4 is actively
communicating with the Central Station.
Minimum QRS Amplitude:
160µV
Recovery from
Defibrillation Discharge
Less than 5 seconds, compliant with IEC 60601-2-25 and/or IEC 60601-2-27.
Summary of Contents for Surveyor S4
Page 4: ...ii TECHNICAL SUPPORT AND SERVICE...
Page 8: ...NOTICES 2...
Page 10: ...WARRANTY INFORMATION 4...
Page 18: ...USER SAFETY INFORMATION 12...
Page 20: ...EQUIPMENT SYMBOLS AND MARKINGS 14...
Page 26: ...ELECTROMAGNETIC COMPATIBILITY EMC 20...
Page 32: ...MAINTENANCE AND CLEANING 26...
Page 56: ...CONFORMANCE TESTING 50...