CONFORMANCE TESTING
46
Test Process
Place the appropriate Test Fixture Insert into Test Fixture (TF-0579) as follows. For Gen 1 device use TF-
0580 for the 3,4,5-wire SAM, or TF-0581 for the 10-wire SAM. For Gen 2 device use TF-0633 for the 3,5-
wire SAM, or the TF-0634 for the 10-wire SAM.
Plug the patient cable end on the S4 Hi-Pot Test Fixture (TF-0619 for 10 Lead SAM) or (TF-0612 for 3/4/5
Lead SAM) into the patient input on the UUT, then insert the UUT into TF-0579 and close the fixture.
Connect the
Red
(positive) test lead of the DC Hi-Pot tester to the terminal on (TF-0619 For 10 Lead SAM
) or (TF-0612 For 3/4/5 Lead SAM)
Connect the
Black
(negative) test lead of the DC Hi-Pot tester to the black terminal of TF-0579.
Set the DC Hi-Pot tester to 5 KVDC and apply power to the test setup by pressing the
Test
button.
Verify the current draw remains less than 200uA for a minimum of 1 second. The display of the Hi-Pot
tester will display a PASS result is the test limits are met or a FAIL result if test limits are not met.
Press the
Reset
button on the Hi-Pot Tester,
If the S4 has the SpO2 option, remove TF-0579 from the black terminal, and connect TF-0643 to the black
terminal of the DC Hi-Pot Tester.
Apply power to the test setup by pressing the
Test
button.
Turn the DC Hi-Pot tester off, and wait 5 seconds.
Remove the UUT from the fixture.
Summary of Contents for Surveyor S4
Page 4: ...ii TECHNICAL SUPPORT AND SERVICE...
Page 8: ...NOTICES 2...
Page 10: ...WARRANTY INFORMATION 4...
Page 18: ...USER SAFETY INFORMATION 12...
Page 20: ...EQUIPMENT SYMBOLS AND MARKINGS 14...
Page 26: ...ELECTROMAGNETIC COMPATIBILITY EMC 20...
Page 32: ...MAINTENANCE AND CLEANING 26...
Page 56: ...CONFORMANCE TESTING 50...