Troubleshooting Chart
APPENDIX B MAINTENANCE AND TROUBLESHOOTING
APPENDIX B
B-1
LCD Message
Problem
Correction
BATTERY LOW
Unable to acquire ECG
Charge the battery with AC power.
or unable to print.
CONNECTION FAILED
Unable to transmit or receive ECGs.
Check for correct baud rate, phone number,
and cable connections or site number.
NO ANSWER
Unable to transmit ECG.
Check for correct phone number. Insure
modems and E-SCRIBE are online.
LEADS OFF OR ONE OR
Lead fail.
Indication of
MORE OF THE FOLLOWING:
RL/RA/LA/LL/V1/V2/V3/V4/V5/V6.
RA, LA, LL,
Check limb leads.
V1, V2, V3, V4, V5, V6
Correct faulty lead(s).
TRANSMIT FAILED
Unable to transmit ECG.
Check phone line. Insure site number
is valid. Try again.
UNABLE TO SAVE ECG
No available memory.
Press stop to continue. Transmit or mark
records for deletion in the directory.
ECG data too noisy to store.
Correct noise and try aquisition/storage again.
LEAD FAULT, NO
Lead fail or noisy ECG data.
Correct faulty lead or noise.
ECG CAPTURE
PAPER QUEUE FAULT
Unable to print.
Add paper; manually advance page evenly
passed closure point of writer and close
writer cover and press STOP.