SECTION 3
3-8
Auto-Fill ID
When enabled, the system will automatically populate the demographic fields in the ID screen. When the
patient ID field is manually populated and followed either by selecting enter or
F2
(
▼
) the system
automatically scans the patient directory and if records with the exact ID are found, the existing data is used to
fill some of the demographic fields. The auto-fill feature is designed to only automatically populate last name,
first name, date of birth, age and gender. If no matching records are found, a brief message is displayed and
the user must manually enter the patient’s demographics.
NOTE:
In order to avoid the use of incorrect data, the Auto-Fill ID feature is only supported if
the ID formats are the same between records and if the records belong to the same site.
AC Filter
The ELI 250 removes 60Hz or 50Hz interference. The setting you select depends on the line frequency in your
country. Always use the 60Hz setting in the United States. If the AC interference is present, check to see that
the proper AC filter is selected.
Paper Speed
The writer speed can be configured at 25mm/s or 50mm/s for default ECG printouts. For rhythm
printouts and display, speeds of 5mm/s or 10mm/s are also available. See Section 4 to change speeds for
display or rhythm printing.
Filter
The ECG plot frequency filter (or print filter) can be set to 40Hz, 150Hz, or 300Hz. The plot frequency
filter does not filter the acquired digital record. A 40Hz plot filter setting will reduce the noise (40Hz and
higher frequencies) on the printed ECG, a 150Hz plot filter setting will reduce the noise (150Hz and
higher frequencies) on the printout and a 300Hz plot filter setting will not filter the printed ECG.
Height/Weight Units
This option defines the units of weight and height to either pounds/inches (lb/in) or kilograms/centimeters (kg/cm).
Date Format
This option defines the format for entering and displaying the patient’s date of birth in either the U.S. format
or the European format.
For example:
July 16th, 2003 is displayed as follows:
US Date Format:
MM/DD/YYYY (07/16/2003)
European Date Format:
DD.MM.YYYY (16.07.2003)
NOTE:
the date format option does not modify the acquisition date printed on each ECG.
Using the example above, the acquisition date would print as 16-Jul-2003.