918S‑‑1130 PP     rev 12‑21‑2020

e. Perform step 2 above to verify the battery level and overall operation of the Alarm.

4. If the “BATTERY LOW” indicator illuminates at any time, replace the internal battery by performing the 

following steps:

CAUTION: 

The following steps should be followed being sure that there is no possibility of fluid ingress 

into the alarm enclosure. 

ATTENTION: 

Les étapes ci-dessous doivent être suivies avec l’assurance qu’il n’y a aucune possibilité

d’entrée de fl uide dans l’enceinte de l’alarme.

Steps

a. Carefully open the 918 Series Alarm enclosure by loosening the Phillips head fasteners at the four 

corners of the cover.

b. Turn off the power switch, SW5 (see Figure 1).

c. Disconnect input device(s) from the Alarm Circuit Board by unplugging the channel connectors from 

the Alarm Circuit Board.

d. Disconnect the battery cable from the Alarm Circuit Board (J7). Gently depress the retaining tab of the 

connector and pull the connector free (see Figure 6).

e. Remove and retain the two screws holding the battery assembly to the rear cover and remove the 

battery assembly from the Alarm (see Figure 7).

f. Install and secure the new battery assembly into the rear cover using the two screws retained in the 

previous step.

f. Install and secure the new battery assembly into the rear cover using the two screws retained in the 

previous step. 

g. Connect the new Battery Assembly to the Alarm Circuit Board’s J7 being sure to fully seat the 

connector. Gently tug on the cable to ensure the retaining tab has clicked into place.

h. Turn on the power switch, SW5 (see Figure 1).  

NOTE: 

If the 918 Series Alarm has been configured to operate with Normally Closed Contacts, the 

Alarm should go into the alarm state when the power switch is turned on; silence this by pressing 

the “Test/Cancel” button.

i. Test the Alarm by pressing and holding the “Test/Cancel” button until the alarm sounds at least three 

times while observing the Channel Alarm and BATTERY LOW indicators (see Figure 1).

j. The 918 Series Alarm indicates the following: 

• The BATTERY LOW indicator will flash momentarily at the beginning of the cycle.

• If the Audible Alarm is configured to sound with an alarm condition, it is loud and strong.

• The Channel Alarm indicators will illuminate when the Audible Alarm is not sounding. If 

configured for no audible alarm, only the Channel Alarm indicator(s) will flash.

• BATTERY LOW indicator is NOT illuminated after the initial momentary flash.

k. If alarm does not sound, the Channel Alarm Indicators does not illuminate or the BATTERY LOW 

indicator illuminates, replace the Alarm Box or call Morrison Bros. Co. Customer Service.

l. Connect the input device to the Alarm Circuit Board. Connect input device to the Channel Input 

Connector (see Figure 2).  

m. Reattach the front cover of the Alarm using the Phillips head fasteners at the four corners of the cover.  

DO NOT OVERTIGHTEN.  “Snug” the fasteners in place and verify that the cover is fully seated to 

the enclosure base.

Figure 6 Disconnecting the Battery from the Control Board

n. It is recommended to simulate an alarm condition and manually trigger the alarm using input device 

(Clock Gauge, float switch or other dry contact device).  If it does not, check wiring at junction box and 

Retaining 

Tab

Depress Retaining 

Tab and pull gently

Summary of Contents for 918D Series

Page 1: ...mer à l un ou l autre des avertissements ou à l une ou l autre des directives apparaissant dans ce document pourrait donner lieu à des déversements de liquides dangereux lesquels pourraient engendrer des dommages matériels des risques de contamination environnementale d incendie ou d explosion des blessures graves ou la mort Specifications The 918S 918D and 918Q are intrinsically safe devices for ...

Page 2: ...te unité autres que celles indiquées dans ces directives engendreront l annulation de la garantie du produit Pour assurer votre sécurité il est important de vous conformer à la réglementation locale d État fédérale ou OSHA régissant les travaux à l intérieur au dessus ou autour du réservoir de stockage et de la zone de canalisation Utilisez tout l équipement de protection individuelle exigé pour t...

Page 3: ...x will be mounted Do NOT attach the wires to the wires in the junction box of the input device s at this time Although not required it is recommended to run the wires in some type of conduit in order to protect them against possible damage and environmental conditions It may be necessary to remove the junction box cover of the input device s INPUT DEVICE WIRING WIRE SIZE LIMITS Input device wiring...

Page 4: ...ill does not sound or the BATTERY LOW indicator illuminates replace the Alarm Box or call Morrison Bros Co Customer Service e Turn off the power switch SW5 before proceeding see Figure 1 4 Connect the two wires from each input device to the Input of the Alarm a Connect the two wires from each input device to their respective Channel Input Connector s see Figure 2 b Verify that the wires are secure...

Page 5: ...ED 918S 918D 918Q 2 Auto Silence ON default ENABLED OFF DISABLED 918S 918D 918Q 3 Channel 1 Normally Closed Open ON Normally Closed Contacts OFF default Normally Open Contacts 918S 918D 918Q 4 Channel 2 Normally Closed Open ON Normally Closed Contacts OFF default Normally Open Contacts 918D 918Q 5 Channel 3 Normally Closed Open ON Normally Closed Contacts OFF default Normally Open Contacts 918Q 6 ...

Page 6: ...sabled the Alarm will flash its Channel Alarm Indicator s and sound the Audible Alarm if so configured until the battery can no longer power the Alarm Box Operational verification should be checked on a more frequent schedule if the Autosilence feature is disabled c CONTACT MODE OPTION i Normally Open Contact Input Devices default If the input device s to be connected to the 918 Series Alarm utili...

Page 7: ...is fitted with this alarm Failure to follow any or all of the warnings and instructions in this document could result in a hazardous liquid spill which could result in property damage environmental contamination fire explosion serious injury or death Le fait de ne pas se conformer à l un ou l autre des avertissements ou à l une ou l autre des directives apparaissant dans ce document pourrait donne...

Page 8: ... Autosilence feature is NOT enabled and the alarm is not silenced by the operator then the 918 Series Alarm will continue to alarm until the battery can no longer power the device 3 To check for an alarm state once the Alarm has silenced itself press and hold the Test Cancel button until the alarm sounds at least three times while observing the Channel Alarm Indicators a If there is an interstitia...

Page 9: ...maintenance operations MONTHLY Test the battery level and overall operation of the Alarm see 2 below YEARLY Check the Enclosure for water ingress see 3 below Battery replacement instructions are in 4 below 1 If necessary clean the Alarm Box with a damp cloth 2 Test the battery level and overall operation of the Alarm no less than once per month a Press and hold the Test Cancel button until the ala...

Page 10: ...clicked into place h Turn on the power switch SW5 see Figure 1 NOTE If the 918 Series Alarm has been configured to operate with Normally Closed Contacts the Alarm should go into the alarm state when the power switch is turned on silence this by pressing the Test Cancel button i Test the Alarm by pressing and holding the Test Cancel button until the alarm sounds at least three times while observing...

Page 11: ...us liquid spill which could result in property damage environmental contamination fire explosion serious injury or death Le fait de ne pas se conformer à l un ou l autre des avertissements ou à l une ou l autre des directives apparaissant dans ce document pourrait donner lieu à des déversements de liquides dangereux lesquels pourraient engendrer des dommages matériels des risques de contamination ...

Page 12: ...12 Morrison Bros Co Dubuque IA 800 553 4840 918S 1130 PP rev 12 21 2020 ...

Page 13: ...13 Morrison Bros Co Dubuque IA 800 553 4840 918S 1130 PP rev 12 21 2020 ...

Reviews: