background image

2

1

2

1

24 Months

< Remove >

< Insert >

• Always insert burs and points all the way in and give them a light tug to make sure they will not come out. 

Otherwise, a bur or point could come out and injure the patient or be swallowed.

• Wait for the handpiece to completely stop rotating before inserting or removing a bur or point.

3. Inserting and Removing Burs and Points

• Protect your fi ngers from being cut by wearing gloves whenever you put in or take out a bur or point.

1

2

Hold the push button down fi rmly and pull the bur or point straight out.

Operation 2/2 

Take out the bur or point and disconnect the handpiece from the motor.

*  Refer to the user manual for the motor.

*  It may be hard to remove a bur or point if it is left in the handpiece.

4. After Use

5. Cleaning

If a medical agent being used for the treatment has adhered to 

the handpiece, wash it off in running water.
Do not perform ultra sonic cleaning.

Wipe the handpiece with a piece of gauze that has been dampened with 

Ethanol for Disinfection (ethanol 70 vol% to 80 vol%) and wrung out 

thoroughly. Clean the end of the light guide if the light seems to be dim.

(only model with a light)

Do not use anything except ethanol. Do not use too much ethanol; it could seep inside and damage the handpiece.
Rotation and spray delivery will be impaired if debris gets inside the water or air lines.
Do not immerse the handpiece in or wipe it with any of the following: functional water (acidic electrolyzed water, 

strong alkaline solution, and ozone water), medical agents (glutaral, etc.), or any other special types of water or 

commercial cleaning liquids. Such liquids may result in metal corrosion and adhesion of the residual medical agent to 

the handpiece.

6. Disinfection

7. Lubrication

Lubricate with the LS Spray (sold separately) or MORITA MULTI SPRAY (sold separately) before autoclav-

ing and after morning and afternoon treatment sessions.

Insert the nozzle into the connection end of the handpiece, 

hold it fi rmly and then spray for 2 seconds.

 Put the spray nozzle on the spray can.

Failure to lubricate the handpiece will result in a malfunction.
Never use any spray other than the LS SPRAY or MORITA MULTI SPRAY. 

The head could get quite cold if the spray is applied for more than 5 seconds. In this case, wait awhile 

before operating the handpiece; otherwise the bearings could be damaged.
Always use the spray can in an upright position.  Use only the special spray nozzle provided.

8. Removing Excess Oil

Connect the handpiece to the motor, point the head 

down, and run the motor for about 30 seconds.  Use a 

piece of gauze to wipe off any oil that gets on the outside 

of the handpiece

.

• After lubricating the handpiece, do not fail to remove excess oil.  The motor could malfunc-

tion if any oil get inside of it.  Also excess oil inside the handpiece could cause it to overheat.

When wiping off excess oil from the head, make sure that you do not touch the push button while 

the handpiece is running.  Doing so will result in wear to the cap and cartridge, making insertion and 

removal of the bur diffi cult.

10.   Troubleshooting

 

Cannot insert burs or points

Take the handpiece off the motor and step on the foot pedal. Does water come out of the motor joint?

If water does not come out

Refer to the accompany user manual for the motor.

If water comes out

Check the o-ring on the motor joint.  Replace it if it is worn or damaged and 

coat it with oil. If the o-ring is OK, clean out the air and water holes on the 

handpiece.  Also take out the water fi lter and clean it. (only model with a 

light : See diagram at right.)  If the spray still does not work, have the hand-

piece repaired.

 

Poor water delivery

 

Poor rotation

Lubricate the handpiece, and then use a new bur or point to check if the problem has been resolved. 

If this does not resolve the problem, have the handpiece repaired.

Before inspecting or adjusting the handpiece, make sure the main power switch is turned on and that the main air and 

water valves are open. If the handpiece does not work properly even after inspection and adjustment, contact your local 

dealer or J. MORITA OFFICE and have it repaired.

Clean drilling and polishing debris off the head with a brush and 

running water.

Clean the spray hole with the spray hole wire provided.

After cleaning, use the syringe or an air blow gun to get rid of any 

moisture left inside the handpiece.

 

Overheating

TERMS AND CONDITIONS

J. MORITA MFG. CORP.

 warrants the Product as follows:

1. 

Warranty Term.

J. MORITA MFG. CORP. warrants that the Product will be free from defects in materials and workmanship for 24 months. 

2. 

Items Not Covered by Warranty.

This Warranty will not cover parts of the Product or consumables which normally need to be replaced such as glass or rubber components, 

and the like. Further, this warranty will not cover defects or damage caused by:
a

Failure to operate or use the Product in accordance with the Operation Instructions or warning/cautionary labels.

b

Modifi cations of the Product, connecting the Product to other products, or using components or replacement parts not specifi ed 

or approved by J. MORITA MFG. CORP. 

c.  Repairs performed by parties not designated by J. MORITA MFG. CORP.

d.  Accidents or forces outside of J. MORITA MFG. CORP. ’s control, such as dropping the Product or collision of the Product with 

other objects, fi re, fl ood, natural disaster, etc.

e.  Use or operation of the Product in settings or under conditions not specifi ed by J. MORITA MFG. CORP. or set forth in the 

Operation Instructions including, without limitation, unauthorized changes in equipment settings, connection to unauthorized 

devices. 

3. 

Remedies of Customer.

Customer must present this Warranty when claiming coverage or application thereof, and either ship or deliver the Product to an 

authorized location of J. MORITA MFG. CORP. or its subsidiaries or authorized dealers.  Upon receipt and verifi cation of this Warranty, 

inspection of the Product, and determination of coverage by this Warranty, J. MORITA MFG. CORP. will, at its sole option, supply the 

parts or components either replaced or repaired free of charge to the Customer. Delivery and shipping charges are the responsibility 

of the Customer. 

4. 

LIMITATIONS AND EXCLUSIONS FROM WARRANTY.

The remedy set forth in paragraph 3 is the sole and exclusive remedy of Customer.  J. MORITA MFG. CORP. MAKES NO OTHER 

EXPRESS OR IMPLIED WARRANTIES AND SPECIFICALLY EXCLUDES ALL OTHER WARRANTIES INCLUDING, WITHOUT LIMITA-

TION, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL J. 

MORITA MFG. CORP. BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL, DIRECT OR INDIRECT, OR SPECIAL DAMAGES 

OF ANY KIND INCLUDING, WITHOUT LIMITATION, DAMAGES FOR LOSS OF REVENUE, LOSS OF BUSINESS OR BUSINESS 

OPPORTUNITY, OR OTHER SIMILAR FINANCIAL LOSS ARISING OUT OF THE PERFORMANCE, USE OR INTERRUPTED USE OF 

THE PRODUCT. 

5.

 

Repairs Not Covered by Warranty.

With respect to repairs not within the period of this Warranty or otherwise not covered by this Warranty, J. MORITA MFG. CORP. will 

perform the same, timely, effi ciently, and for an appropriate cost.  

• To prevent the spread of serious, life-threatening infections such as HIV and hepatitis B, the handpiece must be 

autoclaved after fi nishing the treatment of each patient.

• Instruments are extremely hot after autoclaving; do not touch them until they cool off.

9. Packing and Sterilization

Place the handpiece in a sterilization pouch and autoclave it.

After autoclaving, store the handpiece in a dry environment.

*  Autoclaving and drying temperatures must not 135° C 

(+275° F)

*  Minimum drying time after sterilization: 10 minutes.

Half-moon Groove

End of bur or 

point shaft

Insert the bur up to the half-moon groove inside the chuck. 

Push the bur or point further in until 

you hear it click into place.

Push Button

Make sure fl ows for cooling air and spray are up to specifi cations. Also make sure the motor is operating properly.

Spray Filter

2018. 05.    2,000        Printed in Japan

K

Unit Name

Mode

Detergent

(concentration)

Neutralizer*

(concentration)

Rinse

(concentration)

Miele G7881

Vario TD

neodisher MediClean

(0.3 – 0.5%)

neodisher Z

(0.1 – 0.2%)

neodisher Mielclear

(0.02 – 0.04%)

*  After cleaning there may be streaks or white spots on the instrument.  

Use a neutralizer only if there are streaks or white spots.

*  When using a high-temperature washer-disinfector to clean the handpiece, strictly adhere to the 

conditions specifi ed below.

High-temperature cleaning conditions

Operating conditions for high-temperature washer-disinfectors

Operating Precautions

• 

Always use a handpiece holder when washing the handpiece, making sure to rinse the inside of the handpiece 

thoroughly.

• 

For details on handling medical agents or adjusting their concentration, refer to the user manual for the washing device.

• 

After washing is complete, check to see if the handpiece, including its inside, is completely dry. If any water 

remains inside the handpiece, expel it with an air gun or another such tool. Failure to do so could result in the 

remaining water coming out during use and cause poor lubrication or sterilization.

• 

Always lubricate the handpiece after washing.

Always use a handpiece holder when washing the handpiece; otherwise any medical agent remains inside the 

handpiece, it may corrode or clog the air or water pipe, resulting in a malfunction of the handpiece.

Inappropriate cleaning methods and solutions will damage the handpiece.

Do not clean the handpiece using strong acidic or alkaline solutions that could cause the metal to corrode.

Any moisture left inside the handpiece could result in the remaining water coming out during use and cause 

corrosion, or poor lubrication and sterilization.
Do not leave a handpiece inside the high-temperature washer-disinfector.

• Do not point the spray can at anyone.

• Do not use the spray in a closed space without proper ventilation.  Open windows and turn on 

exhaust fans to make sure the room is well ventilated.

• Do not get the spray in your eyes.  If you accidentally get it in your eyes, wash your eyes with 

plenty of water and see a doctor.

• Do not get the spray on your skin.  If it gets on your skin, wash it off with soap and water.

• If used in a sink or other enclosed space, a fl ammable mist could permeate the air; keep the area 

free of open fl ames or other ignition sources.

Using an air blow gun, blow the handpiece connection for 

20 to 30 seconds to get rid of any moisture left inside the 

handpiece. Wipe the excess oil from the handpiece with a 

gauze.

You can use MORITA's threeway syringe instead of an air 

blow gun.

(In this case, attach the air nozzle to the tip of the threeway 

syringe. The air nozzle comes with a micromotor or its 

optional accessory.)

Recommended Temperature and Time:

In a sterilization pouch, minimum 6 minutes at +134°C (+273.2°F) or 

minimum 60 minutes at +121°C (+249.8°F)

Do not sterilize the handpiece by any method other than autoclaving.
Remove the bur or point before autoclaving.
Failure to clean and lubricate the handpiece before autoclaving will lead to a malfunction of the handpiece.
The sterilization and drying temperatures must not 135°C (+275°F). A drying temperature in excess of 

+135°C (+275°F) will result in contamination of the handpiece or cause it to malfunction.

When autoclaving the handpiece, do not let it come into contact with the heat source or the body of the autoclave to 

avoid damage to the O-rings or other parts due to the high temperature.  
Do not leave the handpiece in the autoclave.

Make sure the motor is properly connected to its tube. Remove the handpiece from the motor and step on the foot 

control. Does the motor run?
If motor runs

Make sure the handpiece is connected to the motor correctly, then step on the foot control to see the bur runs. If 

the bur does not run, have the handpiece repaired.

If motor does not run

Refer to the accompany user manual for the motor.

2018-05-21  Pub.: L102-62001-501

Reviews: