DIAGNOSTICS
ALARM STATUS LIGHT CONDITION
ALL MOTIONS WORK WELL
RE-POWER THE EQUIPMENT
Watch the STATUS lights
Do all three lights, POWER,
TROUBLE and ALARM
illuminate for 5 seconds when you
first power the equipment?
YES NO
There may be a loose or defective
bulb in the socket. The bulb that
isn’t illuminating is the culprit.
Remove the lens cap and correct
the situation which may include
bulb replacement. Look in the
OPERATION MANUAL.
CHECK AC MAINS for
under or over voltage.
AC O.K. = Possible wiring fault
in the BASE. This may also
indicate a failed RENTEC or
FAIRCHILD monitoring module.
Replacement of the MONITOR
STRIP may be necessary, the
assembly that carries both of these
modules.
Gain access to the BASE wiring.
Check all wires to and from the
RENTEC and FAIRCHILD
monitors to be secure.
Morgan MEDesign
, Inc. - Maintenance ~ Service Manual BASIC-ONE -
2005 Page 7
Gain access to the BASE wiring.
Check all wires to and from the
RENTEC and FAIRCHILD
monitors to be secure.
This may also indicate a failed
RENTEC or FAIRCHILD
monitoring module. Replacement
of the MONITOR STRIP may be
necessary, the assembly that
carries both of these modules.