![Morgan MEDesign BASIC-ONE CSA-UL Maintenance And Service Manual Download Page 4](http://html.mh-extra.com/html/morgan/medesign-basic-one-csa-ul/medesign-basic-one-csa-ul_maintenance-and-service-manual_1833851004.webp)
MAINTENANCE SCHEDULE
EL HORARIO DE MANTANIMIENTO
RECEIVED
RECIBIDO
+ 1 YEAR
+ 2 YEARS
+ 3 YEARS
+ 4 YEARS
+ 5 YEARS
___-200__
___-200__
___-20___
___-20___
___-20___
___-20___
ADHERENCE TO ACCESS PROHIBITED NOTICES
Control electronics are provided in a sealed enclosure. It is imperative that only
AUTHORIZED PERSONS
familiar with the working voltages and safety regulations of such
devices attempt to access the control console interior. The following label warning must be
observed and any unauthorized access immediately cancels the warranty.
OPTIONAL MAINTENANCE SCHEDULE AND INSPECTION
When your table was received we recommend that you immediately date the optional
Maintenance Schedule that may have been provided on the base. Optionally use the chart
below or make a copy to keep with your table. It is the responsibility of the user to have annual
inspections and cleaning of the table to continue the validation of the warranty. This process
includes a visual inspection of all moving components and joints and the cleaning of all patient
contact surfaces. Examine the soft control tubes leading to the hand controller for kinks or
cracks. Insure that all electrical cords are not worn or frayed. Remove any grease or soil from
surfaces that may be contacted by patient or user. Check that all fasteners
are secure. Make sure all labels are in
place and continue to be visible.
HARNESS TO HAND CONTROL
ELECTRICAL WIRING
FASTENERS
COSMETIC SURFACES
CLEAN MATS, ACCESS AREAS
LABELS IN PLACE
When the inspection is complete initial the
MAINTENANCE SCHEDULE, check the
small box and insert the inspection date.
AUTHORIZED ACCESS ONLY
Morgan MEDesign
, Inc. - Maintenance ~ Service Manual BASIC-ONE -
2005 Page 3