ENGLISH
28
ELECTROMAGNETIC COMPATIBILITY DECLARATION
Guidance and manufacturer’s declaration- Electromagnetic emissions-
This device is intended for use in the electromagnetic environment specified below. The user
of this device should make sure it is used in such an environment.
Emission test
Compliance
Electromagnetic environment- guidance
Rf emissions CISPR 11
Group 1
The device uses RF Energy only for its internal fun-
ction. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment
RF emissions CISPR11
Class B
The device is suitable for use in all establishment.
Including domestic establishments and those di-
rectly connected to the public low-voltage power
supply network.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies
Guidance and manufacturer’s declaration- Electromagnetic immunity
The Supra devices are intended for use in the electromagnetic environment specified below.
The user of this device should make sure it is used in such an environment.
Immunity Test
IEC60601
test level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic
Discharge (ESD)
IEC61000-4-2
±
6kV contact
±
8kV air
±
6kV contact
±
8kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidi-
ty should be at least 30 %.
Electrical fast
transient/ burst
IEC61000-4-4
±
2kV for power
supply line
±
1kV for
input/out line
±
2kV for
power
supply line
±
1kV for
input/out line
Mains power quality should
be that of atypical
commercial or
hospital environment
Surge
IEC61000-4-5
±
1 kV line(s) to
line(s)
±
2 kV line(s)
to earth
±
1 kV line(s) to
line(s)
Mains power quality should be
that of atypical commercial
or hospital environment.