TROUBLESHOOTING
The information contained in this section covers common problems and suggests solutions.
If your system’s problem is not covered here, or you still need help, please contact the
agent/manufacturer.
Note. Disconnect equipment from the
QPXFSTVQQMZ
electrical supply before
attempting to implement any solutions of an electrical nature
Possible Cause
System does not start.
Check main plug fuse.
Check power supply fuses.
Check all electrical connectors are
mated correctly.
Bad connection.
Solution
Symptom
1PXFS
supply indicator
not lit.
Fuse blown.
Fuses blow repeatedly.
Incorrect supply voltage.
$IFDLFRVJQNFOUSBUFEWPMUBHFJT
DPNQUBCJMFXJUIUIFFMFDUSJDBMTVQQMZ
Disconnect
fi
ltration unit connector
from the power supply. If fuses do not
blow when saw unit is switched on
then the
fi
ltration unit is defective.
Filtration unit has been used
to extract liquids.
System shuts down
repeatedly.
Overload indicator is on.
Distorted blade.
Blunt blade.
Incorrect
fi
tted blade.
Discard
CMBEF
immediately.
Index blade so new portion comes in
contact with the workplace.
Check blade is fitted in accordance
with the manufacturer’s instructions.
See page 8.
See cutting hints
System shuts down.
Overload indicator is on.
Wait until yellow overload indicator is
o
ff
(approx 20 seconds) and retry.
Check connections for the saw unit
and
fi
ltration unit are engaged
correctly with the power supply.
Faulty connection.
Excessive pressure on
workpiece.