4
Section 4
OPERATION
The
POWER
,
BACKLIGHT
and
RANGE
switches on the front panel are
push/push
toggles
and need to be depressed for at least a half second for the unit to respond. The
TEST
switch is a
momentary
. Pressing the
POWER
switch once turns power ON and
pressing a second time turns power OFF. Pressing the
TEST
switch performs a system
test and releasing it resumes normal operation.
A. POWER
– Press for ON, press for OFF –performs this function for either AC or
battery operation.
B.
BACKLIGHT
– At power up, in both AC and DC operation the backlight and status
light are on. Subsequent pressing of the backlight button will cycle as follows:
Both Off
Status Light Only
Backlight Only
Backlight and Status Light
C. RANGE
– Toggles between 2 kV/cm and 20 kV/cm range. Normally, start with the
high range and increase sensitivity for better resolution if the field is less than 2
kV/cm.
D. TEST
– In a normally functioning system when this button is pressed, the meter and
output will immediately go to a significantly more negative value than the measured
value and then return to the measured value when the button is released.
E. ZERO
– This 10-turn potentiometer zeros the instrument to accommodate offsets
generated in the probe. Total span is equivalent to about 300 volts when the probe
is immersed in a zero field. It should be set to produce an indicated value of zero
(with the probe in a zero field). See Section 6 for more details.
F. STATUS
– The STATUS light is a tri-color (and OFF) LED.
AC Power,
Unit ON
LED Green
DC Power,
Unit ON
Battery OK
LED Amber
DC Power,
Unit ON
Low Battery
LED
Alternates
Red/Green
When a low battery condition (< 10.5v DC operation) is detected the display
backlight will blink ON and OFF for 5 minutes and then the unit will shut down.
When in AC operation and power OFF, if charge > 100ma the status light will
blink yellow if it is on. Pressing the backlight switch will toggle it ON/OFF.
Summary of Contents for 257D
Page 11: ...9...