Section 3

INSTALLATION

A. Portable Operation:

The Model 257D Portable Electrostatic Fieldmeter is designed primarily for portable 
operation. Since electrostatic field measurements are made with respect to ground 
be sure to connect the ground cable to a known good ground.

B. Power Line Operation:

An AC power adapter is supplied with each Model 257D.  When the adapter is 
plugged into the receptacle on the rear of the instrument and an appropriate power 
source, the instrument will resort to line operation and the battery will automatically 
be charged as necessary.  The ground cable must still be used as in (

A

).

C. Installation of Probe:

The probe is normally connected directly to the instrument via the DB-9 connector 
recessed into the rear panel.
Probe mounting requirements for electrostatic field determinations vary with the 
nature of the desired measurement. In general, it is best to mount the probe 1 cm 
from the material under test – so long as the input signal remains less than 
20kV/cm. The Model 1036F-5 is calibrated for 20kV/cm at 1cm spacing. 
The backside of the probe assembly is electrically isolated from the rest of the 
probe. A pair of #4-40 threaded inserts are provided for attaching the probe to a 
user provided mount.  It is recommended that, if possible, the probe be mounted 
“looking” downward to minimize the possibility of contamination entering the 
aperture.
The probe may also be hand held at its base for survey measurements. At full scale 
reading, the operator should take care when making measurements as the probe 
may arc to the surface under test.

D. Output:

A BNC connector is provided for an analog output.  This output is 2 volts for full-
scale reading (1/10,000

th

 of 20,000 volts).  The output impedance is less than 10

and is capable of driving capacitive loads

E. Grounding:

A standard banana jack is provided at the rear of the instrument for grounding 
purposes.  A wire cable (supplied) must be connected to a known good ground for 
accurate measurements.

Summary of Contents for 257D

Page 1: ...Operators Manual for Monroe Electronics Inc Model 257D Portable Electrostatic Fieldmeter P N 0340176 257D 100 110309...

Page 2: ...Page 4 BACKLIGHT Page 4 RANGE Page 4 TEST Page 4 ZERO Page 4 STATUS Page 4 MAINTENANCE Section 5 Page 5 Precautions Page 5 Cleaning Page 5 Battery Page 5 Calibration Page 5 PROBE INFORMATION Section 6...

Page 3: ...n re chargeable gel type 12 V 1 2 AH 8 hours with backlight on charging system on board 6 hours re charging time DC power Power adapter supplied Input 100 240 VAC 47 to 63 Hz 400mA Output 18 VDC 840mA...

Page 4: ...put for monitoring A standard banana jack and mating cable with clip is provided for grounding purposes The carrying case has storage space for the probe manual and accessories Applications One major...

Page 5: ...o mount the probe 1 cm from the material under test so long as the input signal remains less than 20kV cm The Model 1036F 5 is calibrated for 20kV cm at 1cm spacing The backside of the probe assembly...

Page 6: ...e field is less than 2 kV cm D TEST In a normally functioning system when this button is pressed the meter and output will immediately go to a significantly more negative value than the measured value...

Page 7: ...for information about probe cleaning Battery Need for battery charging is indicated when the STATUS light begins to flash an alternating red green signal The instrument will continue to operate from...

Page 8: ...y standardized probe instrument combination is dependent upon the gradient cap on each probe Thus the full scale sensitivity for any given system can be determined by inspecting the gradient cap on th...

Page 9: ...rallel and separated by a distance d and have side dimensions of at least 5d The ground plate has a hole in its center just large enough to provide clearance around the probe so that the probe does no...

Page 10: ...ndards Association standard ISA S7 3 1975 or be filtered with a filter such as a Koby Junior filter available from Koby Inc 297 Lincoln Street Marlboro Massachusetts 01752 This filter will provide suf...

Page 11: ...9...

Page 12: ...T ELECTROSTATIC MEASUREMENT TECHNIQUES AND THEIR LIMITATIONS Presented at the ELECTRICAL OVERSTRESS EXPOSITION April 24 26 1984 LT 21 R G CUNNINGHAM Eastman Kodak Co Use of Fieldmeters for Charge Meas...

Page 13: ...repairing or replacing the non conforming goods and or defective work and in accordance with the foregoing Monroe Electronics shall not be liable for any other damages either direct or consequential...

Reviews: