SECTION 1: OPERATION OF THE CABINET 

 

Starting the extract fan 

 
The method of starting the extract system may differ depending on the installation and 
specification. 

 

1.  The extract system may have a starter / on/off buttons near the fume cupboard 
2.  The extract system may be linked to the fume cupboard alarm panel 
3.  The extract system may be controlled by a remote system or BMS 

 
Before operating the fume, cupboard ensure that the airflow alarm has been properly calibrated 
(see separate calibration booklet) 
 
On successful starting of the extract fan, the airflow alarm panel should display a green light. If a 
red flashing light and audible beeping is encountered, please close the sash and report to 
laboratory supervisor.  
(Check to see if alarm panel has been calibrated) 
 

Sash Stop (High) 

 
The glass sash can be raised to its maximum safe height of 500mm for access to the fume 
cupboard chamber.  
The sash will automatically stop at this height.  

 
At all times make sure to close the sash fully, thus providing 
maximum protection to the user. If large equipment is to be 
loaded into the fume cupboard chamber, the sash can be 
raised further by pushing the sash restrictor knob upwards, 
once the sash is lowered the sash stop will automatically reset. 
 
Note: when raising the sash above 500mm an audible alarm 
will be heard and a red flashing light will appear on the sash 

high button of the control panel. Once the sash is lowered below 500mm the sash high button 
light will return to green. 
 

 
Sash Stop (Low) 

 
When the glass sash is fully closed, there will be a 50mm gap at the bottom, this is intentional. 
 

Stopping the extract fan 

 
To switch the extract fan off press the fan OFF button on the control panel / or local stop start. 
 
We recommend running the fan several minutes after an experiment has been completed to 
ensure removal of all fumes. 
 
As a matter of safety please ensure that the sash is always closed unless loading or unloading 
equipment. 

 
 
 
 
 
 
 
 
 
 

Summary of Contents for Ductaire TITAN DT1000

Page 1: ...Operating Maintenance Manual www monmouthscientific co uk Ductaire TITAN Ducted Fume Cupboard DT1000 DT1200 DT1800 DT2000 ...

Page 2: ...y form or by any means redistributed or reproduced in entirety or in part in any form for individual or third party use photocopied or stored on any form of electronic retrieval system or linked to any other website without the prior express written permission of Monmouth Scientific Limited For permission requests please write to Monmouth Scientific Limited Units 5 5 Kilnside East Quay Bridgwater ...

Page 3: ...he fume cupboard chamber The sash will automatically stop at this height At all times make sure to close the sash fully thus providing maximum protection to the user If large equipment is to be loaded into the fume cupboard chamber the sash can be raised further by pushing the sash restrictor knob upwards once the sash is lowered the sash stop will automatically reset Note when raising the sash ab...

Page 4: ...arm panel user manual for further details Light The light switch is located on the left hand service panel at the top press to turn light on and off Double Sockets A switched double electrical socket is usually located on each service panel Services Depending on the fume cupboard configuration various service valves and taps may be fitted which may include cold water and gas etc Turn the valves an...

Page 5: ...h high Leakages Internal seals around panels and sinks etc Check glass sash for cracks chemical attack or any other damage Hygiene Services Illumination Noise Electrical isolation Routine maintenance Lubricate sash cables and pulleys Rocol 34235 Dry PTFE Spray Remove rear baffle and clean inner chamber with mild disinfectant Clean glass sash with standard glass cleaner Fan and Extract System Exami...

Page 6: ...d the exhaust take off section Removal of Front Service Panel To remove the front service panels remove socket head bolts and carefully pull service panel forward this will give you access to the service valves plumbing electrical sockets light switch and the alarm panel Please note Even if the fume cupboard junction box has been isolated the alarm panel can still be electrically connected via ano...

Page 7: ...r from date of delivery or installation with an extension of up to 5 years Subject to T C s Spare Parts Please contact our office for advice on replacement parts Phone 44 0 01278 458090 Option 2 Fax 44 0 01278 458091 Services monmouthscientific co uk ...

Page 8: ...8 Internal Wiring Schematic ...

Page 9: ...0mm W x 2400 H x 900 D Chamber 316g Stainless Steel Work base 316g Stainless Steel Manufacturer Connection Services NONE Drainage NONE Storage NONE Power 2 x D S S O Light LED Alarm Type AFA 1000E VAV alarm panel and controller c w VAV actuator and damper Extract System None ...

Page 10: ...400mm Fume Cupboard Type Ductaire Titan Fume Cupboard Ref XXXX Fume cupboard width 1000mm Sash width 700mm Sash height 500mm recommended 0 00 0 00 0 00 X X 0 00 0 00 0 00 X X 0 00 0 00 0 00 X X 100mm 100mm 100mm 100mm 500mm 700mm Anemometer Anemometer Serial Calibration Date Tested By Test Date Average Velocity Reading Notes ...

Page 11: ...t i s h D e s i g n Manufacture Monmouth Scientific Limited UK Headquarters Units 5 6 Kilnside East Quay Bridgwater Somerset TA6 4DB 44 0 1278 458090 www monmouthscientific co uk info monmouthscientific co uk ...

Reviews: