Monark Ergomedic 839 E
9
Connection to controler
4.2 Connection to an external ECG
device, digital control
Connect the handheld unit to the main unit on
7.
the front of the bike (use the 0-modem cable
with a 9-pin D-sub female at each end).
Connect the net adaptor to a suitable wall
8.
outlet and to the Monark Ergometer 839 and
then turn power on.
After a short while the main menu is shown
9.
on the handheld display.
Press ’99’ and the service menu comes up on
10.
the display.
Press ’6’ for Service setup.
11.
Press ’Enter’ on the following settings until
12.
’Command Type’ is shown.
This says:
13.
- 0 Terminal/PC
- 1-3 see manual
- (0) _
If a zero (0) is displayed at the 3rd line, the
bicycle is in normal mode which means that
the bike can not be controlled by an external
ECG device. Press ‘1’ if a Siemens Megacart
with Ergomed 940 will be used. For use of
other devices press ’2’ or ’3’ and then ’Enter’.
If correct figure is displayed from the begin-
ning just press ‘Enter’.
Return to main menu by pressing ’0’ and
14.
press ’0’ again to finish. The computer will
then save the settings in the memory before
it turns off. Computer will start up again
automatically after a few seconds.
Turn off the power and remove the handheld
15.
unit.
Connect the cable (Megacart needs a specific
16.
cable) into the ECG device.’
The Ergometer can now be controlled only from
an external ECG device.
Reset of the Ergometer to normal mode (control-
led from a handheld unit).
Follow steps 1 – 7. At step 7 press ’0’ and then
’Enter’. The Ergometer can now be controlled
again from the handheld unit or an exetrnal PC.
4.3 Printer
The 839 Electronic Ergometer interfaces to seve-
ral optional devices. A parallel interfaced printer
may be attached to provide written reports.
The terminal or the computer attaches both via a
serial cable to the 9 pin interface connector loca-
ted on the front of the ergometer. Attach a printer
if available to the parallel port on the front.
If the instrument is a terminal or printer, the sys-
tem may need to be setup. Verify that the System
setup have been set to enable automatic printout.
If it has been disabled, no output will reach the
device until it has been enabled. Also, the baud
rate selected by the interface cable must match
that of the device. If the device is a printer, proper
paper loading and unit selection must be comple-
ted prior to operation (refer to printer instruction
manual).
The automatic printout length is a preset to eleven
inch pages for standard fanfold or zee-fold paper.
At the top of the each page, a header designating
the columns is printed.
The time period between the printing of each line
may be set as desired, from 0 (continuous output)
to 255 seconds in one second increments. The
standard setting is 15 seconds between printouts.
This provides reasonable documentation while
not wasting large quantities of paper.
4.4 Analog control
Please see chapter 7.5 Menu 4: Analog control.
Summary of Contents for 839 E
Page 1: ...Manual English Monark 839 E Ergomedic 839 E...
Page 2: ......
Page 17: ...Monark Ergomedic 839 E 17 Computer Reference 7 1 6 Menu system 3 4...
Page 18: ...Monark Ergomedic 839 E 18 Computer Reference Kommandotyp 0 Terminal PC 1 3 Se manual 0...
Page 30: ...Monark Ergomedic 839 E 30 Command type 0 Terminal PC 1 3 Se manual 0 Computer Reference...
Page 40: ...S 780 50 Vansbro Sweden Tel 46 0 281 59 49 40 Fax 46 0 281 719 81 www monarkexercise se...