Monark Ergomedic 839 E
39
10.4 References
Åstrand I, ”Aerobic work capacity in men and women with special reference to age”, Acta Phy-
1.
siol Scand. 49 (suppl. 169), 1960.
Åstrand P-O, ”Experimental studies of physical working capacity in relation to sex
2.
and age”,
Munksgaard, Köpenhamn, 1952.
Åstrand P-O, Rodahl K, ”Textbook ofWork Physiology”, McGraw-Hill, New York,
3.
1970.
Bruce RA, Kusumi F, Hosmer D, ”Maximal oxygen intake and nomographic as
4.
sessment of fun-
ctional aerobic impairment in cardiovascular disease”, Am Heart J 85:546-562,1973.
Naughton J, ”Exercise Testing and Exercise Training in Coronary Heart Disease”,
5.
Academic
Press, New York, 1973.
Golding LA. Myers CR, Sinning WE, ”The Y´s way to physical fitness” YMCAof
6.
the USA, Ro-
semont, IL, 1982.
Wilson PK, Bell CW, Norton AC, ”Rehabilitation of the heart and lungs”, Beckman
7.
instruments,
1980.
Åstrand P-O, ”Ergometri - konditionsprov”, Monark, Sverige.
8.
10.3 Error messages
Message
Reason
Test Aborted
An automatic protocol operation has been
stopped prematurely. No results are available.
Appendix
Symptom
Probable Cause/Corrective Action
Unable to calibrate force
Potentiometer belt may be slipping or broken. Replace if
damaged.
Potentiometer misadjusted.
Reboot memory from service menu (99). Set default (3).
Uneven Heart rate
Use an external unit for example a pulse watch to check
if it also indicates irregular pulse. If it is the case there are
probably disturbance in the room. Magnetic fields from
high voltage cables, elevators, fluorescent tube etc can
cause the disturbance. Other electronic equipment could
be placed to close.
If irregular pulse remains we recommend measuring HR
manually. If HR still remains irregular at workload test
person’s health need to be examined.
Summary of Contents for 839 E
Page 1: ...Manual English Monark 839 E Ergomedic 839 E...
Page 2: ......
Page 17: ...Monark Ergomedic 839 E 17 Computer Reference 7 1 6 Menu system 3 4...
Page 18: ...Monark Ergomedic 839 E 18 Computer Reference Kommandotyp 0 Terminal PC 1 3 Se manual 0...
Page 30: ...Monark Ergomedic 839 E 30 Command type 0 Terminal PC 1 3 Se manual 0 Computer Reference...
Page 40: ...S 780 50 Vansbro Sweden Tel 46 0 281 59 49 40 Fax 46 0 281 719 81 www monarkexercise se...