background image

 

 

pulse) or low (negative pulse).  The user can select either a positive or a 
negative output pulse using the Polarity switch (5). 
The  gain  of  the  input  amplifier  and  the  sensitivity  can  be  adjusted        
manually  or  automatically.    This  allows  the  unit  to  be  used  with  many 
types of targets at various distances and contrasting color conditions.  To 
select  the  Auto  mode,  turn  the  Sensitivity  Adjust  knob  (7)  fully          
counterclockwise, otherwise you are in the Manual mode.  Reflective tape 
and high contrast color applications should be able to use the Auto mode. 

5.1 

Auto Mode 

Make  sure  the  Sensitivity  Adjust  knob 
(7)  is  fully  counterclockwise  and  the 

Auto/Gain

 LED (3) is on.  Aim the laser 

at your target.  Press and hold the Learn/
Gain  button  (6)  until  the 

On  Target

 

LED  (1)  blinks  regularly  or  is  on  solid 
(depending  on  the  RPM  of  the  target).  
Release  the  Learn/Gain  button.    The 

Auto/Gain

 LED will blink to show what 

gain has been selected.  Refer to Table 2. 
In  many  applications,  this  is  all  that  is 
needed.    In  cases  where  the  target  is 
marginal, the user may need to manually 
adjust  the  sensitivity  after  trying  the 
Auto mode. 

5.2 

Manual Mode 

Aim  the  laser  at  your  target.    Turn  the  Sensitivity  Adjust  knob  (7) 
between 

Min

 and 

Max

 until the 

On  Target

 LED (1) starts blinking 

or comes on solid.  If you can‟t get a good setting, change the gain.  

Number of blinks  Gain 

1  Strong Signal 

X 1 

X 2 

X 4 

X 5 

X 8 

X 10 

X 16 

8   Weak Signal -                                   

Move Closer 

X 32 

Table 2  Auto/Gain 

Blinks 

Summary of Contents for SLS-115

Page 1: ...SLS 115 230 Smart Laser Sensor 15 Columbia Drive Amherst NH 03031 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com Website www monarchinstrument com...

Page 2: ...specifications stated This instrument is not user serviceable For technical assistance contact the sales organization from which you purchased the product WARNING Use only the PR Universal charger su...

Page 3: ...0 0 SPECIFICATIONS 9 10 1 Laser Specifications 9 10 2 General 10 10 3 Installation Environment 12 10 4 Compliance 12 11 0 OPTIONS AND ACCESSORIES BACK Monarch Instrument s Limited Warranty applies See...

Page 4: ...Compatible pulse per revolution and has the ability to compute the RPM internally and output the ASCII values to any equipment capable of receiving an RS232 input For best performance use reflective...

Page 5: ...nks RED when fast charging for very low batteries Blinks AMBER when slow charging Steady AMBER when the batteries are fully charged 5 Polarity switch Selects either a positive or negative output pulse...

Page 6: ...her RPMs the LED will be on solid 3 5 If after completing step 3 4 the On Target LED 1 blinks erratically or not at all rotate the Sensitivity Adjust knob 7 slowly from 1 to 9 until the LED blinks reg...

Page 7: ...put A variety of optional cables are available including tinned wire termination 1 8 inch 3 5 mm mono plug or DB9 for RS 232 Refer to Table 1 and Figure 3 for the output socket connection detail DC Po...

Page 8: ...AC charger power supply To turn the unit off slide the Power switch 8 to the Off Charge position If the charger is plugged in the batteries will charge The sensor emits a laser beam which is reflected...

Page 9: ...se and the Auto Gain LED 3 is on Aim the laser at your target Press and hold the Learn Gain button 6 until the On Target LED 1 blinks regularly or is on solid depending on the RPM of the target Releas...

Page 10: ...black target When the unit is first turned on the On Target LED 1 will indicate which target type the sensor is set up for While the Laser On LED 2 is blinking the On Target LED will be on for a white...

Page 11: ...enhanced at long distances by slipping a piece of black tube with a minimum inner diameter of 0 6 inches over the nose piece to act as an extension nose piece This tube should not deflect the beam in...

Page 12: ...supplied PR Universal charger with the sensor 10 0 SPECIFICATIONS Note Product specifications are subject to change without notice 10 1 Laser Specifications Classification Class 3R per IEC 60825 1 Edi...

Page 13: ...tors LEDs for On Target Laser On Auto Gain Power Charge Modes Normal manual or Auto and charging Normal mode manual adjustment of sensitivity gain Auto mode auto learn automatic gain control Sensitivi...

Page 14: ...esistor to user supplied power max 24 V required Polarity Positive or Negative pulse switch selectable RS232 Transmit Out ASCII value of RPM Accuracy 0 002 of reading Settings 9600 baud 8 bits 1 stop...

Page 15: ...32 to 104 F 0 to 40 C operating 14 to 158 F 10 to 70 C storage Humidity Maximum relative humidity of 80 for temperatures up to 88 F 31 C decreasing linearly to 50 relative humidity at 104 F 40 C 10 4...

Page 16: ...h 13 mm wide T 5WP Waterproof Reflective Tape honeycomb pattern 5 foot 1 5 m roll 1 inch 25 mm wide SLS CA BNC Cable assembly with BNC connector SO SLS CA W Cable assembly with tinned leads SLS CA P C...

Reviews: