Nova-Strobe Battery Replacement Instructions
For all Nova-Strobe models
[A]
[B]
[C]
[D1]
[D2]
Dismantling the unit is the easy part. . .
1. Remove the lens [A] by separating the two tabs
that hold it in place.
1a) For LED models, there are two (2)
internal wires connecting the white
connector that will need to be carefully
removed.
2.
Slide the reflector [B] and lamp out as detailed
in the Lamp Replacement section of the User
Manual.
3. The four (4) screws [C] must be removed so the
reflector housing can be detached from the main
body.
4. Remove the four (4) additional screws [D1] & [D2]
in the case halves opposite the input and output
jacks.
5. The case halves can now be separated, exposing
the batteries; the batteries should be sent to a
recycling center or returned to Monarch for
proper disposal.
15 Columbia Drive, Amherst NH 03031-2334 Tel.: 603-883-3390 • Fax: 603-886-3300
www.monarchinstrument.com [email protected]
Before you open the housing,
know that during the reassembly there are multiple points
that need to align correctly to close the housing securely.
Some of them include:
A —
Display board sits in groove
A —
DC Connector spans both case halves
B —
Mounting T-nut sits in slot as shown
B —
Lanyard knot surrounds screw standoff and protrudes through the notch in the case
C —
Trigger sits in case brackets with wires tucked out of the way
D —
PCB aligns in a retaining groove
D —
Battery holders (2) nests in brackets, there is a right and wrong way to insert them