Molift Quick Raiser 205 User Manual Download Page 12

11

Molift QuickRaiser 205 

Transport and operating conditions

Transport and Storage

For long time storage it is recommended that the 

emergency stop button is activated (pushed in). 

The lifter can be stored and transported under 

temperatures between -25 - 70 °C.

Operating

The lifter is designed for use at standard room 

temperatures (+5 to +40°C).

Air Pressure: 70 - 106 kPa

Relative Humidity: 15 - 93 %

Storage or transport at other temperatures leave 

the lifter in a room with a suitable temperature 

until it reaches a safe operating temperature.

Medical electrical equipment 

requires special precautions regard-

ing electromagnetic compatibility 

(EMC). Portable or mobile radio 

communication equipment may 

affect the medical electrical equip-

ment, and should be kept minimum 

25 cm (10 inches) from the lifters 

electronics.

Battery

Molift Quick Raiser 205 is available with a 24V SLA 

battery or a 25.6V LiFePO4 battery - both with an 

integrated charger. 

The SLA battery has a life expectancy of approx. 

500 charge cycles. 

The LiFePO4 battery has a life expectancy of 

approx. 2000 charge cycles.

Batteries must be disposed of as 

special waste according to local 

rules and regulations. Do not 

dispose in general waste

Battery and Service indicator

Wrench symbol / 

Service indicator
Battery symbol / 

Battery indicator

The electrical system has a power-save function

which will turn off the electrical system five

minutes after last operation. All indicators will turn 

off.The system is activated when pushing one of 

the operating buttons

Battery indicator on control panel

When the multi level battery indicator is yellow 

the lifter will have sufficient power available for 

one full lifting cycle with max load. When battery 

is flashing yellow it is only possible to lower the 

lifting arm.

Battery indicator

Battery level

Green

80% or more

Green

80 - 60%

Green

60 - 40%

Green

40 - 20%

Yellow or 

yellow 

blinking

20% or less

(recharge lifter)

Electronics

Summary of Contents for Quick Raiser 205

Page 1: ...Molift Quick Raiser 205 EN User manual BM13001 Rev C 2018 01 08...

Page 2: ...leg support 13 Using a 4 point lifting arm 14 Using a 2 point lifting arm 14 Raising 15 Transfer 15 Accessories 16 Accessories 16 Suspensions and Slings 16 Maintenance 17 Cleaning and disinfeciton 17...

Page 3: ...rs and an electric motor powered by a battery The hoist is operated by using the hand control or the control panel on the hoist It has electrically adjustable legs and with a lifting capacity of 205 k...

Page 4: ...es class 1 and has been tested and approved by a third party according to standards IEC 60601 1 IEC 60601 1 2 and NS EN ISO 10535 2006 Conditions for Use Lift and transfer of a person will always pose...

Page 5: ...Applied parts Sling Foot rest Rocker arm handles Technical Data Quick Raiser 205 922 mm 3619 64in 808 mm 3157 64in 574 mm 2219 32in 530 mm 2055 64in 1068 mm 423 64in 1320 mm 5131 32in 1730 mm 687 64i...

Page 6: ...All screws with pre applied threadlock should only be used once Spare part art no 2920075 1 Battery Push in the emergency stop button before assembly Place the column on a horizontal surface Make sure...

Page 7: ...sis Mount the screw with washer on column Slide the screw down to chassis and tighten the screw Turn the lifter over on the side Insert screw and washer in the top hole and tighten the screw Connect c...

Page 8: ...ert the bolt and tighten the screw using the 17mm spanner and an 6mm allen key Make sure the lifting arm is properly fastened to the trolley on the column Fit the cover onto the trolley 5 Mains cable...

Page 9: ...ns Make sure the leg support and lifting arm are properly fixed and that there are no loose parts on the lifter Make sure that the lifting column is properly fixed to the chassis with both screws tigh...

Page 10: ...inimum 18 minutes Duty Cycle 10 Intermittence according to standard ISO EN 10535 Safe Working Load The lifter is marked with Safe Working Load SWL If maximum load SWL differs between lifter suspension...

Page 11: ...ush up or kick on top part of the brake to unlock Emergency stop Emergency lowering Emergency Stop The emergency stop shuts off the power when pushed in Turn clockwise to reset Electrical emergency lo...

Page 12: ...24V SLA battery or a 25 6V LiFePO4 battery both with an integrated charger The SLA battery has a life expectancy of approx 500 charge cycles The LiFePO4 battery has a life expectancy of approx 2000 c...

Page 13: ...odic inspection required Red Perform service Charging LED The lifter has an integrated charger Park the lifter and connect to a wall socket to charge the battery A LED light on the back of the battery...

Page 14: ...turers Adjusting leg support Move the Quick Raiser 205 in front of the user Place the feet on the footplate If necessary help the user to place the feet correctly Lock the wheels on the lifter before...

Page 15: ...lightly when lifting and if possible hold onto the handles Removing sling Loosen the straps again by angling the buckle and puling the strap out Using a 2 point suspension On the 2 point suspension ar...

Page 16: ...y fastened 5 Lift user to standing postition and tighten and adjust the sling if necessary Make sure the users hands are placed correctly to avoid that the lifting straps are squeezing the users hands...

Page 17: ...ns Art no 2910059 Set of small cushions Art no 2910076 Lifting arms and slings 4 point suspension Quick Raiser Active lifting arm Art no 2910060 Rgosling Active Art no 1720910 Size XS Art no 1720920 S...

Page 18: ...able on request Periodic Inspection Scope Periodic inspection is a visual examination particu larly of the lifter s load bearing structure and lifting mechanism with attachments brakes controls safety...

Page 19: ...Short circuit between the hand control and lifter clean the connectors of the hand control and chassis with spirits to remove grease and dirt The hand control has failed exchange hand control contact...

Page 20: ......