Chapter 18: Maintenance
Perform only the maintenance tasks described in this guide. Contact a Molecular Devices
service engineer to inspect and perform a preventive maintenance service on the instrument
each year. See
Before operating the instrument or performing maintenance operations, make sure you are
familiar with the safety information in this guide. See
.
Maintenance and troubleshooting procedures that can be done by users to ensure optimal
operation of the instrument are described as follows.
Doing Preventive Maintenance on page 227
Cleaning the Instrument on page 228
Testing the Stackers on page 232
Testing the Pins on page 233
Aligning the Camera on page 235
Replacing Fuses on page 237
Draining the Compressor on page 238
Moving the Instrument on page 239
Adding a New User to the SQL Server Database on page 240
Running the Software as an Administrator on page 240
Resetting the Path to the SQL Server on page 240
Troubleshooting on page 243
CAUTION! Maintenance procedures other than those specified in this guide must be
performed by Molecular Devices. When service is required, contact Molecular Devices
technical support.
Doing Preventive Maintenance
You are responsible for doing daily and weekly maintenance. Also, Molecular Devices strongly
recommends that a complete instrument maintenance be done every six (6) months by an
approved service engineer.
Daily Maintenance
Ensure that the interior of the instrument is free from dirt and dust. Check the surface of the
x-drive drag-chain support bracket. Dust and debris collected here can be swept onto the
instrument bed during operation. See
Cleaning the Instrument on page 228
Before storing the head, or whenever it needs cleaning, wipe the exterior of the head with a
cloth using 70% ethanol. To store the head, slide it into its metal cover with the pins facing
inward and then secure it in place with the thumbscrew and washer.
Remove, clean, and sanitize the wash bath and brushes.
Check the air regulator for signs of moisture. If necessary, twist the drain clockwise
approximately a half turn to force out moisture. Some types of regulators might require the
drain purge to be pushed up towards the regulator bowl.
18
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...