Chapter 19: Troubleshooting
This section describes some common problems that can occur and possible solutions.
Generally, crash recovery involves some form of the following procedure:
1. An error is identified in the software or the instrument stops.
2. If the error condition is caused by a receptacle or other item, then remove the item from the
instrument.
3. Press the red reset button on the end of the stacker lane with the error.
4. Reactivate the software, generally by clicking Next.
Common Problems and Possible Solutions
The instrument does not start. The system does not function correctly
Make sure that the main power switch is turned on at the wall, the Emergency Stop button is
pulled out, and the Start button is pressed. See
Pre-Power-Up Check List on page 27
Make sure that the power cord is seated fully in the mains input of the instrument. See
Instrument Connections on page 19
Make sure that the mains input fuse is in good condition. See
Turn off the instrument power and then turn it on again.
The computer does not start
Make sure that the power switch on the computer is in the on position and that the main
power switch is turned on at the wall.
Make sure that the fuses are in good condition. See
One or more of the axes do not move
Make sure that the main power switch is turned on at the wall, the Emergency Stop button is
pulled out, and the Start button is pressed. See
Pre-Power-Up Check List on page 27
The drive system fails to home the actuator
Make sure that the door is closed.
Manually move the head to the center of the instrument then close the door and retry the
process. See the Knowledge Base answer number
.
The UV light does not turn on
Make sure that the door is closed.
Make sure that filaments in the lamp are not burned out.
Lamp failure
Determine whether the halogen lamps must be replaced.
The Picking alignment is incorrect
Inspect the head for loose or bent pins.
Run a Pin Fire Test. See
.
The pins bend when inoculating in plate wells
Make sure that the plate is aligned correctly towards both arrows in the plate holder.
Make sure that the correct plate, head, and spacer blocks are installed.
19
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...