monolayer are recommended to ensure better mixing of
compounds and lower signal variance across the plate. However,
further assay development, adjustment of the volume, height
and speed of dispense, is recommended to optimize your cell
response.
Recommended Setup Parameters
Recommended experimental setup parameters are as follows.
Experimental Setup
Parameters
EMCCD
Cameras
ICCD Camera
Read Mode
Fluorescence
Fluorescence
Excitation Wavelength
470–495 nm
470–495 nm
Emission Wavelength
515–575 nm
515–575 nm
Excitation Intensity
80%
50
1
%
1.
Can be adjusted to increase or decrease the signal if low RFUs or saturation
problems occur.
Camera Gain
50–130
2000
Gate Open
—
6
1
%
Exposure Length
0.4
1
s
0.53
1
s
Read Interval
1 s
1 s
Addition
Parameters
96-Well
Plate
384-Well
Plate
1536-Well
Plate
Compound
concentration (fold)
5X
5X
7X
Addition volume (
μ
L)
50
12.5
1
Addition height (
μ
L)
210–230
25–45
1–2
Tip up speed (mm/s)
10
10
5
Dispense speed (
μ
L/s)
Adherent cells
50–100
30–40
4–7
Dispense speed (
μ
L/s)
Non-adherent cells
10–20
10–20
1–5
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...