Exchanging Hardware
146
0112-0109 H
during instrument operation but may require attention to be
avoided when lifting a plate by hand.
Cell Reservoir
Installing Cell Reservoir
Initial installation has to be performed by a Molecular Devices certified
field service engineer, however if the Cell Reservoir was removed
following instructions in
Uninstalling Cell Reservoir. on page 147
,
reinstall it as follows:
1.
Ensure the tubing connection to the connector is tight.
2.
Align the
Reservoir Connector
to the Cell Reservoir and
tighten the two
Captive Thumb Screws
.
3.
Insert the Cell Reservoir into the Source Plate 3 position.
4.
Go to the
Instrument
menu and select
Refresh
Configuration
. The
Instrument Status
tab displays
Cell
Reservoir: OK
.
This status indicates that only the Cell Reservoir can be selected
for the Source Plate 3 position.
5.
Perform a
Manual Wash
(see
Wash Cell Reservoir Process on
) and inspect the cell reservoir area to ensure that
there is no fluid leakage.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...