Uninstalling Cell Reservoir.
When the Cell Reservoir is configured and detected by the instrument,
it will be the only available choice as the Source Plate 3 selection.
ScreenWorks displays
Cell Reservoir: OK
in the Instrument Status
tab. In order to use the Source 3 position for a different source plate,
the Cell Reservoir will have to be uninstalled.
1.
Remove the Reservoir Connector by loosening both Captive
Thumb Screws.
2.
Place the Reservoir Connector by the Tip Washer.
3.
Remove the Cell Reservoir.
4.
Click
Instrument > Refresh Configuration
.
The
Instrument Status
tab displays
Cell Reservoir:
NOT_OK
.
This status indicates that another plate can now be selected for
the Source Plate 3 position.
Note:
The Cell Reservoir can be autoclaved at this time.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...