FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
117
Mix Fluid Process
Fluid mixing steps can be added to protocols as independent protocol
steps—
Mix Fluid
—or linked to a
Transfer Fluid
step—
Mix with TF
.
Independent Mix Fluid steps can be configured to mix fluid in source or
read plates, while
Mix with TF
steps immediately follow the Transfer
Fluid step they are linked to, mixing fluid in the read plate where
compound was dispensed.
Different mix options are offered for standard pipettors and the pin tool
(described in different sections below).
Mix Fluid (Standard Pipettor)
Fluid mixing with a standard pipettor consists of a series of aspirations
and dispenses in and out of the tips.
Read
Check this box to link a
Read
process to the
Transfer
Fluid
process.
This automatically adds a
Read with TF
step to the
protocol, and opens the configuration page for that
step.
See
about read process options.
Note:
This option is required if you intend to
read a plate while transferring liquid during a
kinetic cell-based assay (for example, calcium
mobilization).
Mix Fluid
Check this box to link a
Mix Fluid
process to the
Transfer Fluid
process.
This automatically adds a
Mix with TF
step to the
protocol, and opens the configuration page for that
step.
about mixing options.
Unload Pin Tool Automatically unload the pin tool after the dispense.
Item
Description
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...