3.5 Configuring Instrument Settings in Flex Mode
3.5.4 Automix
The Automix function is a patented feature that allows you to set automated shaking of
the microplate before and/or in between readings.
Select Before First Read to shake the plate before the first wavelength reading only. You
can also set Automix to shake the plate Between Reads.
•
Click in the box next to the type of shaking you want.
•
Type a value in the box on the right to indicate how long shaking should last.
Automix time can last from 1 to 999 seconds.
Automix Settings
3.5.5 AutoCalibrate
This checkbox allows you to disable or enable automatic calibration. The default is
enabled. Turn autocalibration off to allow the instrument to begin or complete readings
more quickly.
Important!
Allow the instrument to perform an autocalibration of at least one plate
before you disable this function.
Note:
The instrument maintains the most recent autocalibration settings in memory
(NVRAM) until another autocalibration is performed.
AutoCalibrate Selection
FlexStation II Operator’s Manual – Rev. D
59