Troubleshooting
Common Problems
Problem
Possible Solutions
Machine will not turn on
-
Check that main power switch is turned on and
the emergency stop button is released (pulled
out)
Computer will not start
-
Check that PC power is turned on
One or more of the axes won’t move
-
Check that the main power is turned on and
the START button has been pressed
-
Manually move the actuator to the middle of
the bed
-
Disconnect and reconnect control cable
between PC and right side of ClonePix
2 (a
small screwdriver may be necessary).
-
Re-boot ClonePix
2
Interior light or UV germicidal lamp will not turn on
-
Check that the bulbs are not burned out
Picking alignment incorrect
-
Ensure that pins are correctly aligned Check
for any bent pins and replace if necessary
Silicon tubing becomes detached from pin during
purge
-
Pin may be blocked. Replace with new pin
Water is leaking from picking head or actuator
-
Pin may be blocked. Replace with new pin
Poor picking results
-
Air in system. Purge the supply line
-
Check Z height of pin
Not depositing
-
Pin blocked – run Purge option or replace pin
-
Dispense volume too low
No liquid flow to Wash Station
-
System fluid supply bottle is empty
-
Check air supply and liquid pipes for kinks
-
Check outputs in configuration are correct
(consult Customer Support)
Wash Station overflowing
-
Check no kinks in tubing to waste bottle
-
Waste bottle full and/or vacuum pump broken
Pins are catching on holes of the wash station
-
Datum point not set accurately.
-
Pins are damaged
-
Pin fitted with 2 O-rings
Water is leaking from top of pins
-
Check each pin is fitted with one o-ring
Pins not firing correctly
-
Check that air compressor is working
-
Pistons may have accumulated residue. Run
Pin Fire Test for up to 1 hour. Do not undertake
a Pick Run if pins not firing correctly
Illumination cover not opening closing correctly
-
Check that air compressor is working
No ethanol flow to Replenishing Ethanol Wash Bath
-
Check no kinks in ethanol supply tubing
-
Peristaltic pump tubing has become blocked
(usually after period of inactivity)
Replenishing Ethanol Wash Bath overflowing
-
Check no kinks in ethanol waste tubing
-
Peristaltic pump tubing has become blocked
(usually after period of inactivity)