16
Chapter 2. Connections and controls
GAIN IN
Voltage-controlled proportional gain of fast servo,
±
1
V
, correspon-
ding to the full-range of the front-panel knob. Replaces front-panel
FAST GAIN
control when
DIP1
is enabled.
SWEEP IN
External ramp input allows for arbitrary frequency scanning, 0 to
2.5 V. Signal must cross 1.25 V, which defines the centre of the sweep
and the approximate lock point.
MOD IN
High-bandwidth modulation input, added directly to fast output,
±
1
V
. Requires
DIP4
to be enabled
1
.
SLOW OUT
Slow control signal output, 0 V to 2.5 V. Normally connected to a
piezo driver or other slow actuator.
FAST OUT
Fast control signal output,
±
2
.
5 V. Normally connected to diode in-
jection current, acousto- or electro-optic modulator, or other fast
actuator.
MONITOR 1, 2
Selected signal output for monitoring.
TRIG
Low to high
TTL
output at sweep centre.
LOCK IN
TTL
scan/lock control; 3.5 mm stereo connector, left/right (pins 2, 3)
for slow/fast lock; low (ground) is active (enable lock). Front-panel
scan/lock switch must be on
SCAN
for
LOCK IN
to have effect. Digikey
cable
CP-2207-ND
provides a 3.5 mm plug with wire ends; red for slow
lock, thin black for fast lock, and thick black for ground.
1
2
3
1
Ground
2
Fast lock
3
Slow lock
Figure 2.2:
3.5 mm stereo connector pinout for
TTL
scan/lock control.
1
When DIP4 is enabled, MOD IN must be terminated when not in use.