DANSENSOR
68
User Guide
EN
CheckPoint 3
P/N 340486-B
09/2017
4. Now the calibration is saved...
...and the following message appears.
5. Tap
OK
to confirm and return to the
Calibration
start screen.
Calibrate: 100% CO2
NOTE! Before calibrating with 100% CO
2
you must calibrate with 0% CO
2
.
1. Select
Calibrate: 100% CO2
.
Now follow procedure as described for the 0% CO
2
calibration on
page 67
, making sure to
expose the needle to a 100% CO
2
gas.
Restore defaults
Resets device to factory calibration settings.