![MMR Technologies elan2 digital User Manual Download Page 15](http://html1.mh-extra.com/html/mmr-technologies/elan2-digital/elan2-digital_user-manual_1813982015.webp)
DATASHEET & ELECTRICAL
Voltage:
115V 60Hz, Single Phase
Power Consumption (RUN):
~950W
Power Consumption (STANDBY):
10W
Noise dB (RUN):
~65dB (1 meter)
Noise dB (STANDBY):
Quiet
Air Source:
90-120PSI (-55C°/ 6ppm Dew Point)
Liquid Nitrogen Production:
Up to 8Liters per day
(depending on environment)
Liquid Nitrogen Purity:
N2 99%
Residual O2 / Argon 0.4%
Liquefier Weight:
72.2 lbs (32.8kgs)
Liquefier Dimensions:
13(L) x 13(W) x 32(H) inches
Air Compressor Weight:
45 lbs (26.1kgs)
Air Compressor Dimensions:
17(L) x 17(W) x 15(H) inches
ELECTRICAL:
This product is intended for indoor use only and must be connected to a proper
source of power as designated by the label at the rear of the system. Use only the cables supplied
with the device and ensure a proper Ground is present. Irregular / unstable power may lead to
hardware damage. Use of equipment other than specifically intended may result in permanent
damage to the system, or impairment of the safety devices.
RISK OF ELECTRIC SHOCK IF OPENED OR SYSTEM IMPROPERLY GROUNDED.
REFRIGERANT:
The core refrigeration system of the Elan2D Liquefier contains a multi-component
proprietary refrigerant mix, allowing Liquid Nitrogen temperatures to be possible. Although some
individual components in the mixture are flammable the overall mixture is considered inert.
DISPOSAL:
Under US EPA and International Laws governing the safe handling and disposal of
refrigerants the contents of the system must be recovered by a qualified refrigeration technician in
accordance with local environmental regulations.
PAGE 13