TROUBLESHOOTING
ERROR STATE:
When an Error condition is encountered a code is displayed and the system is
shutdown to prevent damage, the cooling fans and air compressor will remain on. While the
systems have been designed and calibrated to prevent false alarms this will happen occasionally
when certain monitored parameters such as temperature or humidity go out of range sporadically
/ momentarily.
ERROR CODES:
The list of possible Error codes and actions.
ER.00
Low Production Time-out Error. Three (3) consecutive attempts and defrosts
have been made to produce within a given time (5.5hrs) have failed.
Action: Contact MMR.
ER.01
Hardware Failure – Flow Sensor Error.
Action: Contact MMR.
ER.02
High Humidity Error.
Action: Check room humidity and temperature. Service Air Compressor.
ER.03
Hardware Failure – Liquefier Level Sensor Error.
Action: Contact MMR.
ER.04
High Power Consumption (>1500W).
Action: Leave unit in STANDBY for 24 hours, resume with a DEFROST.
Contact MMR.
ER.05
Low Air Pressure Error (<60PSI for 30 seconds).
Action: Check Air Hose for leaks. Service Air Compressor.
ER.06
Cool Down Time-out Error. Three (3) consecutive attempts and defrosts have
been made to cool down within a given time (2.5hrs) have failed.
Action: Contact MMR.
ER.07
Compressor High Temperature Error (>90C°).
Action: Check for adequate ventilation, check room temperature, replace rear
filter, check for vent obstruction. (See Maintenance and Care).
ACKNOWLEDGE AND RESTART:
Press and hold “TRASNFER” 7 and “LOCK” 9 buttons
simultaneously until the unit restarts. As a precaution, a 90 minute defrost will begin.