background image

When it is level, secure the feet by tightening the lock nuts 
with a 10 mm spanner. 

4.2 

SHELVES

 

AND

 

USABLE

 

SPACE

Each incubator is supplied with 4 perforated stainless steel 
shelves. The shelves locate into removable stainless steel 
racking. The shelves incorporate an anti-tilt mechanism, 
stopping the shelves from being pulled out too far.

Each shelf measures 690 mm wide by 390 mm deep. The 
minimum spacing between shelves is 80 mm. There are 8 
possible positions for the shelves.

The shelves and racking can be left in place during 
decontamination.

The usable space within the incubator is the volume 
in which the conditions are within our published 
data.

A, B, C = Internal dimensions
a = distance from left and right wall = 70 mm
b = distance from top = 70 mm
c = distance from front and back wall = 70 mm
d = distance from base = 70 mm

For practical purposes this is the perforated area on each 
shelf.

x

  Do not place any samples outside of this space.

x

 'RQRWRYHU¿OOHDFKVKHOI

x

 $OORZVXI¿FLHQWURRPEHWZHHQVDPSOHVWRDOORZIRU

DLUÀRZ

x

  Do not use the top shelf for very deep samples.

4.3 

WATER

 

TRAY

A stainless steel water tray is supplied which allows the 
incubator to provide 90 % rH.

3ODFHWKHWUD\LVRQWKHÀRRURIWKHLQFXEDWRUORFDWLQJLWZLWK

the side racking.

&DUHIXOO\¿OOWKHWUD\ZLWKOLWUHVRIGLVWLOOHGVWHULOLVHG

water.

000UHFRPPHQGWKDW\RXFOHDQDQGUH¿OOWKHWUD\HYHU\

week.

Any spillages within the chamber should be removed and 

FOHDQHGLPPHGLDWHO\:DWHUOHIWRQWKHFKDPEHUÀRRUZLOO

cause condensation to form.

Under no circumstances should any equipment be placed in 

WKHZDWHUWUD\RURQWKHFKDPEHUÀRRU

Remove the water tray before moving the incubator.

Instructions for use

  

6

                      

CO

2

CELL Standard 190_np_en 1301_mmm_V1.02

 

Empty the water tray if the incubator is switched off, 
otherwise condensation will occur.  

4.4 

GAS

 

CONNECTION

 

Risk of injury. Compressed gas.
Risk of damage to unit.

Observe relevant regulations for 
handling CO

2

Risk of asphyxiation. Site must have 
adequate ventilation.

Risk of CO

2

 poisoning. Site must have 

adequate ventilation.

The incubator is designed for use with CO

2

 of  

99.5 % purity. Input pressure greater than 1.0 bar 
(15 psi) will damage the incubator so a pressure 
reducing valve (PRV) must be used.

The CO

2

 inlet to the incubator is via a 4 mm barbed push 

¿WWLQJ
359¶VZLWKWKHFRUUHFW¿WWLQJVDQGRWKHUJDVDFFHVVRULHV

can be supplied by MMM. 

D

Section 8

For correct connection of CO

2

, follow these steps in order:

 &RQQHFWWKHVXSSOLHGWX(3$¿OWHUWRWKHJDV

supply.

2)  Fully close the PRV.
3)  Open the main valve on the gas bottle or central supply.
4)  Adjust the PRV pressure of the gas supply to between 

0.3 bar and 0.7 bar (4.5 to 10 psi).

5)  Close the main valve on the gas bottle / central supply.
6)  Connect the other end of the tubing to the incubator CO

2

 

inlet. Secure with the supplied clip.

Repeat these steps whenever you change the CO

2

 bottle or 

supply.

Turn off the main valve and disconnect from the CO

2

 supply 

if the incubator is to be unused for extended periods.

4.5 

ELECTRICAL

 

CONNECTION

Risk of injury.
Risk of damage to unit. Observe correct 
mains supply voltage.

Risk of electrical shock. Do not connect 
with wet hands.
No user serviceable parts. Do not 
remove service panels.

Disconnect unit in an emergency.

Before connecting the incubator, check that the electrical 
supply corresponds with the values on the incubator data 

SODWH,ILQGRXEWFRQVXOWDTXDOL¿HGHOHFWULFLDQ

The incubator is designed to operate from a single phase 
electrical supply of: 230 V AC ±10 % 50/60 Hz.

This incubator 

MUST

 be connected to a protective earthed 

supply.

Summary of Contents for CO2Cell 190 Standard

Page 1: ...Instructions for use CO2 INCUBATOR CO2Cell 190 Standard CO2Cell 190D Standard with decontamination 160 C CO2 CELL Standard 190_np_en 1301_mmm_V1 02...

Page 2: ......

Page 3: ...4 3 WATER TRAY 6 4 4 GAS CONNECTION 6 4 5 ELECTRICAL CONNECTION 6 5 GETTING STARTED 7 5 1 START UP 7 5 2 FACTORY PRESETS 7 5 3 MAIN DISPLAY 7 5 4 CONTROL MENUMAP 7 5 5 CHANGING TEMPERATURE SET POINT...

Page 4: ...ANING 12 9 1 OUTER SURFACES 12 9 2 INNER SURFACES 12 9 3 OXIDATION AT HIGH TEMPERATURE 12 10 SERVICE 12 10 1 WARRANTY 12 10 2 PREVENTATIVE MAINTENANCE 12 10 3 RETURNING A UNIT 12 11 DISPOSAL 13 12 TEC...

Page 5: ...ctices for working in a laboratory environment These safety instructions are intended to be used alongside any local safety guidelines rules and regulations If in GRXEW SOHDVH FRQVXOW WKH HDOWK DQG 6D...

Page 6: ...ed to the controller alarm system DSection 5 13 An inner glass door seals against a thermoplastic elastomer gasket This allows viewing of the samples within the chamber without disturbing the temperat...

Page 7: ...ITIONS The incubator can be placed on a table or worktop 3OHDVH HQVXUH WKDW WKH VLWH LV VXI FLHQWO VWURQJ enough to carry the weight of the incubator MMM do not recommend placing the incubator directl...

Page 8: ...LL Standard 190_np_en 1301_mmm_V1 02 Empty the water tray if the incubator is switched off otherwise condensation will occur 4 4 GAS CONNECTION Risk of injury Compressed gas Risk of damage to unit Obs...

Page 9: ...temperature alarm 1 0 C relative to set point 2W Low temperature alarm 1 0 C relative to set point F High CO2 alarm 1 0 C relative to set point 2F Low CO2 alarm 1 0 C relative to set point GRG Door op...

Page 10: ...set point which means that they do not need adjusting when you change the chamber temperature The over and under temperature alarms will follow the changes in your chamber temperature For example If...

Page 11: ...for 1 second will show in the left display Press 63W 63F Press for 1 second W 2W F for 1 second W will show in the left display Press If an alarm sounds on start up press to silence it for 1 second 2W...

Page 12: ...amination cycle Leave the CO2 pipe connected Instructions for use 10 CO2 CELL Standard 190_np_en 1301_mmm_V1 02 5 13 DOOR OPEN ALARM 7KH LQFXEDWRU LV WWHG ZLWK D GRRU RSHQ DODUP V VWHP WR alert you wh...

Page 13: ...d the ambient temperature So an incubator placed under a worktop will take longer to cool than a freestanding incubator on a worktop We recommend that you carry out the decontamination cycle at the en...

Page 14: ...e was the incubator located x Contact details of the person who reported the fault x FRPSOHWHG GHFRQWDPLQDWLRQ FHUWL FDWH SOXV D FRS faxed through in advance MMM cannot accept any returns without a co...

Page 15: ...0 s opening minutes 13 CO2 MANAGEMENT Range CO2 0 20 Control CO2 0 1 Stability CO2 0 2 Uniformity CO2 0 2 Door recovery 30 s opening minutes 6 CO2 sensor type NDIR CO2 inlet pressure Bar 0 3 to 0 7 CO...

Page 16: ...Instructions for use 14 CO2 CELL Standard 190_np_en 1301_mmm_V1 02...

Page 17: ...CO2 0 0 7 0 0 8 9 6 6 9 6 Europ melweisstras d precision CO CO2Cell Com 0 6 0 4 6 4 6 8 7 6 6 7 0 6 6 7 pean Directiv sse 6 D 8215 O2 incubator mfort CO2 A B 3 1 8 C B 3 1 8 C 6 7 D 7 E ve 2006 95 EC...

Page 18: ......

Page 19: ......

Page 20: ...M N L O P L Q R S O Q S N T P U O V L O W X Y Z L X X L L S P Q _ L a b c d e f d g _ O L V V W L Q X _ O h i j k l m n m n m m o o p o q r j k l m n m n m m o o p s q t u v w x y z v t t t v v v j v...

Reviews: