2.0 Pre-installation Considerations Continue 

2.3 Utility Requirements 

Coolant 

The InSight X3, X3+, X3+DA is a closed loop coolant system and requires no external utility water 

services. Use ONLY full-strength 

Nalco 

460-PCCL104 liquid corrosion inhibitor as a coolant. 

Never use deionized water. 

Fill the chiller reservoir  following the instructions in the chiller 

manufacturer's user manual. Do not place the chiller above the laser.  Should the unit not be 

installed properly,  and a leak develops, dripping water may damage the laser. 

The chiller must always be on when the power supply is on, even if the diode lasers are not on. 

Power Requirements 

The laser and the chiller must be connected to separate AC power circuits. 

IPS-300 Power Supply  

Electrical Single Phase 

Voltage 100-120VAC/ 200-240VAC, 50/60 Hz 

Current 10A/ 6A max. 

Heat Load is 1.44kW max. at 20C. 

Chiller SSC ThermoRack 401 

Voltage 120-240VAC, 50/60 Hz 

Current 9A max 

Heat Load is 2.16kW max. at 20C. 

2.4 Computer Requirements 

Two modes of control are offered for the standard InSight system: 

A.

InSight graphical user interface (GUI) control software for use on a Windows

©

 base   computer

through the supplied USB cable. For more information on the control software, refer to the

User's Manual.

B.

Serial commands that can be sent to the InSight laser head from the user’s host system using

the supplied RS-232 serial cable.

For more information on the control software, refer to the User's Manual. In order to run the GUI 

control software, the control computer must meet these minimum requirements: 

• Intel or AMD 32 or 64-bit, single or multi-core processor with > 1 GHz clock 

speed

• 1 GB RAM

• 10 MB available disk space for installation

• an USB 2.0 or higher port.

• a mouse or other Windows

®

-compatible pointing device

• a video display with 1024 x 768 (SVGA) or higher resolution

• Microsoft Windows 10

®

 operating system

5

Summary of Contents for Spectra-Physics InSight X3+

Page 1: ...KS Spectra Physics www spectra physics com InSight X3 Systems Pre installation Guide InSight X3 and X3 A laser systems Spectra Physics Lasers 1565 Barber Lane Milpitas CA 95035 408 980 4300 sales spec...

Page 2: ...thin one week of your delivery date you will receive a call from one of our factory trained Customer Service Engineers Some preliminary planning is essential to avoid unnecessary delays during install...

Page 3: ...end as much time as possible reviewing the system components before your Spectra Physics Customer Service Engineer arrives for the installation and training Refer to Chapter 4 of the User s Manual for...

Page 4: ...m operating room temperatures are 20 25 C Room temperature should ideally be 22 C and should not fluctuate more than 1 C during any two hour period See specifications in Table 1 1 and table 1 3 in the...

Page 5: ...00 240VAC 50 60 Hz Current 10A 6A max Heat Load is 1 44kW max at 20C Chiller SSC ThermoRack 401 Voltage 120 240VAC 50 60 Hz Current 9A max Heat Load is 2 16kW max at 20C 2 4 Computer Requirements Two...

Page 6: ...ngineer s arrival unless specifically told to do so by the factory Two or more people will be required to lift some equipment _ Compare the packing list with your quotation Call your Spectra Physics o...