MKS MicroPirani 925 Series Short Form Manual Download Page 4

The analog pressure output (16 bit resolution) is based on the digital 
pressure reading (PR1). Connect the signal to a differential input. 

 

 

Do not connect the negative side of the analog output to 
the power supply return or to any other ground. The 
voltage drop from supply current will produce errors in the 
analog output. 

 
Converting voltage to pressure: 

P = 10

(Vout – c)

 

V

out

 = c + log

10

(P) 

 

 

mbar 

5.5 

Torr 

5.625 

Pascal 

3.5 

 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 

 

Vout vs. Pressure (Torr / mbar, N

2

 setup) 

The transducer can provide other analog output characteristics than the 
standard 1VDC/decade. Refer to Operation and installation manual. 

 
Gas dependence 

The MicroPirani™ transducer is based on measurement of thermal 
conductivity and consequently its reading depends on gas type and gas 
concentration. The transducer is calibrated for Nitrogen, and will read higher 
pressure values when exposed to atmospheric air. 
 
 