Electrical Specifications
Appendix A: Product Specifications
28
Warning
Units with NW-16-KF fittings and a full scale range greater
than 1200 Torr (23 psi) require an HPS overpressure ring.
Operating the unit without a protective overpressure ring may
result in injury.
The HPS part number for the overpressure ring is HPS 10031.
Electrical Specifications
Input Required
0 to 5 Volt output
0 to 10 Volt output
10.8 VDC to +32 VDC (regulated if below 13 VDC)
@ 10 mA max.
+13 VDC to +32 VDC @ 10 mA max.
Output
0 to 5 VDC
0 to 10 VDC
into >10 K ohm load
into >10 K ohm load
Due to continuing research and development activities, these product specifications are subject to change without notice.