Setup
Chapter Two: Installation
10
Fittings
The 722 transducer is available with a variety of different fittings, listed below:
•
½
Tube
•
4-VCR
type Female
•
8-VCR
type Female
•
8-VCO
type Female
•
NW-16-KF
•
Mini-CF
Caution
1. MKS
does not warranty
the 722 transducer when single or double
metal ferrule compression-type vacuum fittings are used because
damage will occur to the transducer when improper tightening
procedures are followed.
2. Before proceeding to
Setup
, page 9, carefully check all plumbing
connections to the instrument to ensure a leak-tight installation.
Making Mechanical Connections
To make mechanical connections in line to a system, use the recommended installation practices,
as specified by the fitting manufacturer or by an appropriate standard.
Warning
Improper installation can cause personal injury or damage
equipment. Follow proper installation procedures at all times.
How To Tighten Fitting Attachments
For VCR Fittings:
Tighten ¹/
8
turn past the finger tight position for 316 SS or nickel gaskets
(¼ turn for copper or aluminum).