MJK Expert™ 3400 Level Transmitter is 

designed for use in waterworks, sewage 

treatment plants and industrial applications 

where accurate and long term stable level 

measuring is required. It is very robust and 

designed for applications in rough envi-

ronments with aggressive chemicals and 

aggressive fluids.
A major design feature is the programming 

facility. With a standard USB interface and 

a laptop computer, the servicing transmit-

ter can be programmed on-the-fly to the 

required measuring range.

Specifications

60 mm

236 mm

Dimensions and 

Accessories

Cable bracket for MJK model 3400
(1 pc. is delivered with each transmitter)

Order numbers

MJK Expert™ 3400 
Level Transmitter

135 mm

32 mm

Dia. 50 mm

1” NPT

Dia. 50 mm

135 mm

36 mm

Data Sheet for 

Expert™ 3400 Level 

Transmitter

Order Numbers

2051

Expert™ 3400, range 0 - 0.3 mWG

2057

Expert™ 3400, range 0 - 100 mWG

2052

Expert™ 3400, range 0 - 1 mWG

2058

Expert™ 3400, range 0 - 300 mWG

205

Expert™ 3400, range 0 - 3 mWG

2061

Expert™ 3400, range 0 - 0.3 m 1" RG top

2054

Expert™ 3400, range 0 - 5 mWG

2062

Expert™ 3400, range 0 - 1 m 1" RG top

2055

Expert™ 3400, range 0 - 10 mWG

206

Expert™ 3400, range 0 - 3 m 1" RG top

2056

Expert™ 3400, range 0 - 30 mWG

20225

Expert™ 3400 w/ special meas. range

MJK Expert™ 3400 Hydrostatic Level Transmitter

Nominal measuring range

0 - 0,3 m

0 - 1 m

0 - 3 m 

0 - 5 m

0 - 10 m

0 - 30 m

0 - 100 m

0 - 300 m

Measuring principle

Ceramic capacitive, relative pressure

Min. programmable range

0 - 0,3 m 0 - 0,3 m

0 - 1 m

0 - 3 m

0 - 3 m

0 - 10 m

0 - 30 m

0 - 90 m

Max. programmable range

0 - 1 m

0 - 1 m

0 - 3 m

0 - 10 m

0 - 10 m

0 - 30 m

0 - 100 m

0 - 300 m

Max. overpressure

4 bar

4 bar

6 bar

10 bar

10 bar

18 bar

40 bar

40 bar

Temperature range

-20 °C to 40 °C

Temp. deviation, zero oint

Better than ± 0.02 % / °C

Temperature deviation, FS

Better than ± 0.01 % / °C

Linearity / Stability

Better than ± 0.2 % FS / ± 0.2 % FS

Measurement accuracy

Better than ± 0.1 % FS @ +10 to 30 °C. Better than ± 0.2 % FS @ full temperature range

Long time stability

Better than ± 0.1 % FS per year

Materials

Housing: stainless steel (AISI 316 L), diaphragm: 96,6 % Al

2

O

3

/Cell packing: Viton

®

Supply voltage

10 - 30 V DC (12 - 30 V DC for cable lengths above 110 m)

Output signal

2-wire 4 - 20 mA (passive transmitter)

Cable

2 × 0,5 mm

 (pressure) + 5 x 0,15 mm

2

 (data), shielded, PUR insulation

Cable length

12 m

12 m

12 m

12 m

12 m

35 m

110 m

320 m

Enclosure

IP 68, withstands a static pressure equal to max. overpressure.

Approvals

ATEX: II 2G EEx ia IIC T5/T6, CE: EN61000-6-3/-4:2001, EN61000-6-1/-2:1999, 
EN50014:1997 E, EN50020:2002

 

 Can be delivered with other cable lengths on request.

Summary of Contents for 202925

Page 1: ...l 45 45 56 06 56 Fax 45 45 56 06 46 mjk mjk com www mjk com As our products are continuously improved we reserve the right to make any change in the stated specifications and dimensions without prior...

Page 2: ...ct complies with the requirements concerning electromagnetic compatibility EMC stipulated in Council directive no 89 336 EEC of 3rd May 1989 altered at directive no 92 31 EEC on the approximation of t...

Page 3: ...air 4 Mechanical mounting 5 Model 1400 5 Model 3400 5 Model 3400 with thread top 5 Electrical mounting 6 Cable length vs supply voltage 6 Designation of wires cutting stripping the cable 6 Explosion h...

Page 4: ...s are printed on a label on the transmitter housing For all versions the pressure range is indicated on the label Safety instructions 1 Read this manual carefully 2 Be aware of the environment on the...

Page 5: ...itting by sliding the two plastic jaws upwards place the cable between the jaws and slide the jaws downwards until the cable locks Secure the cable fitting by pulling the cable downwards 3 Lower the p...

Page 6: ...transmitter or make any attemt to program the tranmitter while the transmitter is located in an explosion hazardous zone Electrical mounting Cable length vs supply voltage The cable can be lengthened...

Page 7: ...r The Z788 Zener Barrier MJK part no 202995 has a safety fuse which will blow if the current exceeds the limit The voltage protection in the Zener barrier contains a number of zener diodes which are c...

Page 8: ...ength Cmax Ci Ccable 129 m Cable length based of the inductance Max cable length Lmax Li Lcable 2862 m In this case the capacitance sets the limit of the cable length to 129 m If another approved isol...

Page 9: ...ur or mists Equipment category Explosive gas air mixtures G Category 1 Zone 0 1 or 2 Category 2 Zone 1 or 2 Category 3 Zone 2 Designation of explosion protection EEx ia IIC T6 Electrical apparatus wit...

Page 10: ...02993 the max cable length is 129 m Using The Z788 Zener Barrier MJK Order no 202995 the max cable length is 99 m Alternatively other cable types with equivalent specifica tions may be used If a cable...

Page 11: ...mechanical parameters 4 Install the device only in media for which the wetted materials have sufficient durability 5 Expert transmitter type 1400 3400 hydrostatics level transmitter may not be used i...

Page 12: ...tter than 0 2 FS 0 2 FS Measurement accuracy Better than 0 1 FS 10 to 30 C Better than 0 2 FS full temperature range Long time stability Better than 0 1 FS per year Materials Housing stainless steel A...

Page 13: ...op 203955 Expert 3400 range 0 10 mWG 203963 Expert 3400 range 0 3 m 1 RG top 203956 Expert 3400 range 0 30 mWG 202925 Expert 3400 w special meas range MJK Expert 3400 Hydrostatic Level Transmitter Nom...

Page 14: ...14 Manual Manual Expert 1400 3400 GB 0704 MJK Automation A S Byageren 7 DK 2850 Naerum Denmark Tel 45 45 56 06 56 Fax 45 45 56 06 46 mjk mjk com www mjk com This page intentionally left blank...

Page 15: ...15 Manual Manual Expert 1400 3400 GB 0704 MJK Automation A S Byageren 7 DK 2850 Naerum Denmark Tel 45 45 56 06 56 Fax 45 45 56 06 46 mjk mjk com www mjk com This page intentionally left blank...

Page 16: ...16 Manual Manual Expert 1400 3400 GB 0704 MJK Automation A S Byageren 7 DK 2850 Naerum Denmark Tel 45 45 56 06 56 Fax 45 45 56 06 46 mjk mjk com www mjk com...

Reviews: