7
2
Safety precaution
CAUTION
●
Devices and accessories used together with this product
•
Before using other devices or accessories, thoroughly read the instruction manual
of the devices and make sure that the operating table is not affected adversely.
Before fitting on accessories from third party companies, contact your distributor or
Mizuho. Some accessories cannot be fitted on.
•
While operating the operating table, check the position of other devices or the
accessories used with them. They may come in contact with each other during the
operation, operating table, devices, and/or accessories may get damaged.
•
For hygiene, be sure to use sterilized drapes on the areas on this product where
the patient comes into contact with it.
●
Cleaning and disinfection
•
After using the operating table, make sure to follow the steps below to clean up
and disinfect the operating table.
1. Turn off the power and disconnect the power cord from the medical grade outlet.
2. Detach all the mattresses from the operating table.
3. Use a lint-free cloth soaked with proper volume of disinfectant to wipe off the
upper, side, and back side of the mattresses.
4. As with step 3, disinfect the surfaces of the tables and side rails.
5. Wipe off the operating table with a clean dry cloth 15 minutes after disinfecting it.
•
Make sure to use Mizuho authorized disinfectants. The disinfectants are as shown below.
a) Sodium hypochlorite 0.1% (halogen containing compound)
b) Hypo Alcohol (iodine color removing agent)
c) Chlorhexidine (chlorhexidine gluconate 0.5%)
d) Benzalkonium chloride (invert soap 10%)
e) Povidone iodine
f) Ethanol 80%
g) Hydrogen peroxide
h) Saline (0.9%)
i) Isopropyl alcohol (IPA)
●
Moving and transporting
•
Do not move the operating table with a patient on it.
•
Follow the procedures below to move the operating table.
* Before moving the operating table, disinfect the entire operating table in order to
prevent infection.
1. Turn off the power and disconnect the power cord from the medical grade outlet.
2.
Check if the handles and levers are in fixed positions, and each section is fixed firmly.
3. Unlock the brakes, and move the operating table.
* Do not allow the operating table to move in conjunction with the head plate or
leg plate when they are being pulled out from the operating table.
•
The operating table should be transported with the following conditions met.
1. Disinfect the entire operating table before transporting it.
2. Take measures to prevent it from tipping over, such as lowering the tabletop to
the bottom position.
3. Actuate the brake.
4. Suitably position cushioning on the product to prevent it from getting damaged
during transport.
5. Store the product in a container so that it does not get exposed to dust, and the weather.
●
Disposal
In accordance with the European Union Waste Electrical and Electronic Equipment
(WEEE) Directive, all electrical components and batteries must be disposed of in
accordance with local regulations. Please contact your local distributor for proper
disposal.