3-2. Cleaning Recommendations
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
The following procedures should be followed when
cleaning the surgical table between cases.
Place table top in level position prior to starting
cleaning procedure.
WARNING
Always follow OSHA blood-borne patho-
gens standards for protective clothing,
including gloves, masks and eye pro-
tection when cleaning the surgical table.
Remove major contaminants from the table with
disposable materials following appropriate biohaz-
ard waste disposal procedures.
Remove all table pads and place them on a flat
surface for cleaning.
CAUTION
Thoroughly read and follow the
manufacturer's directions for all clean-
ing fluids. DO NOT use cleaners con-
taining phenolics.
Apply cleaning fluid liberally to top and sides of
each pad and wipe with a clean lint-free cloth.
Using a clean, damp, lint-free cloth, wipe the pads
to remove the cleaning fluid.
Using a clean, dry, lint-free cloth, wipe the pads to
remove all moisture.
Repeat the steps to clean the bottom of the each
pad.
3-1. Preventive Maintenance
The following preventive maintenance checks and
services are recommended to ensure the service-
ability and proper operation of your Surgical Table,
and should only be performed by qualified trained
personnel.
a. During normal cleaning, a general visual ex-
amination should be made checking for leaks,
loose bolts or parts, and cracked, chipped, or
missing paint. Any necessary repairs should be
made.
b. Semi-annually the following checks and ser-
vices should be performed:
1. Check all hydraulic fittings, mini-valves and
slave cylinders for proper operation and
any signs of leaks.
2. Check the hydraulic speed controls and
adjust if necessary.
3. Pressure check (with a gauge) the pres-
sure relief valve.
4. Check all mechanical adjustments and ad-
just as necessary.
5. Check hydraulic fluid level.
6. Lubricate the slider assembly.
7. Check function of foot leg adjustment le-
vers. Lubricate as necessary.