![Mitsubishi MELSEC-Q Q62HLC-U-HW User Manual Download Page 9](http://html1.mh-extra.com/html/mitsubishi/melsec-q-q62hlc-u-hw/melsec-q-q62hlc-u-hw_user-manual_1808777009.webp)
3
2.2 Types, Measured Ranges and Data Resolutions of Usable Input Sensors
Input
Input range
Digital value
Data resolution
K
-200 to 1372°C
-2000 to 13720
J
-200 to 1200°C
-2000 to 12000
T
-200 to 400°C
-2000 to 4000
S
-50 to 1768°C
-500 to 17680
R
-50 to 1768°C
-500 to 17680
N
0 to 1300°C
0 to 13000
E
-200 to 1000°C
-2000 to 10000
B
0 to 1800°C
0 to 18000
PL II
0 to 1390°C
0 to 13900
Thermocouple
W5Re/W26Re
0 to 2300°C
0 to 23000
0.1°C
0 to 10mV
0.5 V
0 to 100mV
0 to 20000
5 V
-10 to 10mV
1 V
Micro voltage
-100 to 100mV
-10000 to 10000
10 V
0 to 1V
0.05mV
1 to 5V
0.2mV
0 to 5V
0.25mV
0 to 10V
0 to 20000
0.5mV
-1 to 1V
0.1mV
-5 to 5V
0.5mV
Voltage
-10 to 10V
-10000 to 10000
1mV
4 to 20mA
0.8 A
Current
0 to 20mA
0 to 20000
1 A
2.3 Indicated Accuracy for Ambient Temperature
(1) Ambient temperature 23 2°C
Item Error
Less than -100°C
1.0°C
-100 to less than 500°C
0.5°C
K,J,T,E,PL II
500°C or more
(indicated value×(0.1%)+1digit)
-50 to less than 1000°C
1.0°C
S,R,N,
W5Re/W26Re
1000°C or more
(indicated value×(0.1%)+1digit)
Less than 400°C
70.0°C
400 to less than 1000°C
1.0°C
Thermocouple
B
1000°C or more
(indicated value×(0.1%)+1digit)
Micro voltage
Voltage
Current
Full-scale×( 0.1%)