mvip@mitech-ndt.com
Units: Metric and Imperial unit selectable.
Battery information indicates the rest capacity of the battery.
Auto sleep and auto power off function to conserve battery life.
USB2.0 communication port
Bluetooth support.
Size: 120mm×67mm×31mm
Weight: 295g
1.4 Transducer Technical Parameters
Table 1.1 Transducer Technical Parameters
Model Freq
MHz
Φ
mm
Measuring
Range
Lower
limit
Description
N02
2.5
14
3.0mm
~
300.0mm
(
In
Steel
)
40mm (in Gray
Cast Iron
HT200)
20mm
for thick, highly
attenuating, or
highly scattering
materials
N05
5
10
1mm
~
600.0mm
(
In Steel
)
Φ20mm×
3.0mm
Normal
Measurement
N05
/90°
5
10
1mm
~
600.0mm
(
In Steel
)
Φ20mm×
3.0mm
Normal
Measurement
N07
7
6
0.65mm
~
200.0mm
(
In Steel
)
Φ15mm×
2.0mm
For thin pipe wall
or small curvature
pipe wall
measurement
HT5
5
12
1
~
600mm
(
In Steel
)
30mm
For high
temperature
(lower than 300
℃
)
measurement.
P5EE
5
10 P-E: 2~600mm
E-E: 3~100mm
Φ20mm×
3.0mm
Normal
Measurement
and
trough-coating
thickness testing