-44-
151153EN-G
DMC 3000 User's Guide
A
dd-
on Mo
dules
10.4 Module Detachment
For detachment, ensure the dosimeter is in pause Mode.
To clear “MOD dEf” alert indication from dosimeter display, it is recommended to perform the following:
Select “Detach module” in the bottom of DMCUser’s window or remove the dosimeter’s battery and then put it
immediately back into place.
Note :
If the external module configuration is with " Module required" (for example "Neutron required" the Detach Module action
cannot clear the "MOD dEf" alert indication.
Due to current being present on the pins when the DMC 3000 is energized, take care when attaching new modules or adapter
board not to short any of the pins.
The add-on module must be attached or removed only when the DMC 3000 is in PAUSE.