Mira CR4000 Operational Instruction Manual Download Page 8

 

 

 

 

 

 First through innovation

 

 

 

 

0086

 

 

Mira, Inc.*414 Quaker Highway * Uxbridge, MA 01569 * Tel: 508-278-7877 * 800 847-6472 * Fax: 508-278-4555    p.

8

 

                     

3.8 

CR4000 Instructional Video (CR40x)

 

 

The  CR4000  system  instruction  video  is  available  in  several  formats:  CD,  NTSC,  SECAM  or 
PAL. A videotape or CD is supplied with each CR4000 system. The format received is customer 
specified. The video contains an overview of the CR4000 system and is intended to complement 
this manual. It can also be used as an instructional aid. 

 

3.9 

Ophthalmic Probes 

 

 

All probe shafts and tips are stainless steel. Refer to Table  II for a listing of the MIRA 
Ophthalmic  Cryo  Probes.  The  gas  expansion  chamber  is  at  the  very  tip  of  the  probes, 
thereby assuring that freezing and defrosting occurs only at the target area. Warm high-
pressure gas insulates the shaft during freezing and defrosting cycles eliminating the need 
for  and  insulating  boot  while  reducing  the  danger  of  freezing  adjacent  tissues.  The 
handles  present  an  unobstructed  view  of  the  probe  shaft  and  remain  comfortable  to  the 
hand during use.  

TABLE II Probe Model Numbers

 

Model No; 

 

Description   

 

 

Tip Diameter Ice Ball Size 

 

CR4010           Curved Retinal 

 

2.8mm  

 

3.5mm 

CR4015           Straight Retinal 

 

2.8mm  

 

3.5mm 

 

CR4022 

Straight Cataract 

 

 

 

 

2mm    

 

3.0mm 

CR4023           Curved Cataract

 

 

2mm    

 

3.0mm 

CR4025           Mini Curved Cataract 

 

1.5mm  

 

2.0mm 

CR4030           Curved Glaucoma 

 

3.5.mm 

 

4.0mm 

CR4040           Vitreous 

 

1.5.mm 

 

2.5mm 

CR4045           Mini Straight Vitreous 

 

1.5 mm 

 

1.5mm 

CR4075          Curved Hammerhead 

 

2.8mm x 6mm long 

4.5mm  

CR4080          Curved Baby Hammerhead 

 

1.5mm x 4mm long 

2.5mm 

CR4085          Gaydon ROP 

 

2mm   

 

3.0mm 

CR4090          Abramson Retinoblastoma 

 

1.5mm  

 

2.0mm 

 

CR3095 

Lashkari Trichiasis 

 

 

 

 

2.8mm  

 

2.5mm 

 

4.0 

INITIAL SET-UP PROCEDURE

 

 

IMPORTANT: To be performed or double checked each time an element of the CR4000 
system is added, changed or disassembled. 

Obtain  a  non-siphon  gas  cylinder  of  N

2

O  or  CO

2

.  The  type  of  cylinder  and  gas  selected  will 

determine the connector

 

needed.

 

See table 1 for a complete selection of gas cylinder connectors. 

 

WARNING; 

Siphon gas cylinders will vent gas in a liquid state when the valve is opened. 

Do 

not use siphon gas cylinders

 

   

Summary of Contents for CR4000

Page 1: ...First through innovation 0086 Mira Inc 414 Quaker Highway Uxbridge MA 01569 Tel 508 278 7877 800 847 6472 Fax 508 278 4555 p 1 OPHTHALMIC CRYOUNIT CR4000 OPERATIONAL INSTRUCTION MANUAL...

Page 2: ...er Hose CR4160 6 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 7 3 8 CR4000 Instructional Video CR40x 8 3 9 Ophthalmic Probes 8 4 0 Initial Set Up Procedure 8 5 0 Foot Switch Operati...

Page 3: ...t which can obscure the surgeon s view The MIRA CR4000 has a rapid freeze and defrost cycle available providing greater control of application time Temperature settings for both CO2 and N2O allow for...

Page 4: ...or changed When a cylinder is heated above room temperature it may exceed maximum pressure 900psi red zone In this situation allow the cylinder to equilibrate to room temperature Approx 70 F before us...

Page 5: ...o Carriers that come with factory installed sintered metal filters With a Micro fiber or sintered metal Filter assembly attached For use with basic CR4000 systems IMPORTANT Micro fiber or sintered met...

Page 6: ...ures as described in section 6 0 Sintered metal Filter element and O Ring replacement procedure CR5350 Proceed as in items 1 2 and 3 above 4 Remove the filter element 5 Inspect the sealing ring and re...

Page 7: ...r N2O CR4290 T Yoke assembly for E 4in Cylinder CO2 CR4390 T Yoke assembly with Micro filter for E 4in Cylinder CO2 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 The CR4150 is specia...

Page 8: ...ues The handles present an unobstructed view of the probe shaft and remain comfortable to the hand during use TABLE II Probe Model Numbers Model No Description Tip Diameter Ice Ball Size CR4010 Curved...

Page 9: ...er hose to the rear panel fitting labeled System Exhaust Connect the other end to an aspiration system or equivalent Note Use of the scavenger hose will result in quieter operation of the CR4000 Check...

Page 10: ...he 5 C temperature setting for Carbon Dioxide CO2 6 3 4 Depress the foot pedal for 10 seconds and then release 6 3 5 Wait 60 seconds while the exhaust gas is escaping from the rear scavenger port IMPO...

Page 11: ...for repair WARNING Do not alter the shape of the probe tip The shape of the probe tip is vital to its proper function Any change to the probe tip will void the manufacturer s warranty 8 2 Cleaning The...

Page 12: ...s completely non electric there are no electrical shock hazards The micro fiber filter Sintered metal filter assemblies prevent foreign particles or oils from entering the CR4000 console and probes Re...

Page 13: ...efore patient treatment Turn the gas cylinder OFF and depressurize the CR4000 system before changing the Micro filter Sintered metal filter or O ring When the probe is not in use the protective cap sh...

Page 14: ...er Place all probes in plastic bags Enter a brief note that includes the following Return Goods Authorization RGA Your full address for return of shipment Please include specific department Descriptio...

Page 15: ...ncurred by the buyer with the purchase or operation of the equipment or components thereof Repair to damaged parts will be made at a minimal charge The buyer shall pay all shipping charges on the equi...

Reviews: