7.0 PRESSURE REDUCTION REGULATORS

 

 

For optimum performance of the Mini-Vitreous Probe CR4045, a standard pressure reduction regulator 
should be used to reduce the gas pressure delivered to the CR4000 system. A suitable two-stage pressure 
reduction regulator is used instead of the cylinder connector. The gas pressure from the cylinder should 
be adjusted to indicate 625psi on the CR4000 front panel gauge. 
 

A  pressure  reduction  regulator

  is  only  recommended  for  the  Mini-  Vitreous  Probe  CR4045.  For  all 

other probes, the pressure should be set between 680 and 900psi(700psi is preferred), within the green 
zone  of  the  front  panel  gauge.  Contact  your  MIRA  representative  for  more  information  on  pressure 
reduction regulators. 
 

8.0 PROBE CARE 

8.1 

 

Use

 

The  Ophthalmic  Cryo  probes  must  be  handled  with  care.  They  are  non-electric  and  are 
easily inserted or removed from the console’s receptacle. Do not use the probe if the tip is 
scratched or damaged. Return scratched or damaged probes to MIRA for repair. 

 

WARNING;

 Do not alter the shape of the probe tip. The shape of the probe tip is vital to 

its proper function. Any change to the probe tip will void the manufacturer’s warranty. 

 

8.2. 

Cleaning  

 

The probes can be cleaned in accordance with the published protocol. 

The entire probe 

should  never  be  submerged  in  any  solution

.  The  probe  jack  cap  should  always  be 

attached  during  cleaning  or  sterilization  to  prevent  foreign  particles  from  entering  the 
body. 

 

8.3 

Storage 

 

Storing the probes in tight coils is not recommended. Coil loops should be a minimum of 
6 inches. The white caps will protect the probe tip and probe jack when being transported 
or stored. 

 

    

8.4

 

Sterilizing  

 

CR4000 probes are made of stainless steel, Teflon and silicone rubber. The probes should 
be  sterilized  in  accordance  with  hospital  protocol  and  Mira,  Inc.  IFU.  It  is  the  user’s 
responsibility to select the sterilizing conditions that do not harm these materials. These 
materials will tolerate steam sterilization.  

 

WARNING; 

The  probe  jack  cover  must  be  screwed  into  place  before  sterilizing 

regardless of the sterilizing methods used. If moisture from steam or gas enters the probe 
during sterilization, it will adversely affect the performance of the system.  

                       

Please ensure that the probe tip cover is on at all times when the probe is not in use.

          

Summary of Contents for CR4000

Page 1: ...First through innovation 0086 Mira Inc 414 Quaker Highway Uxbridge MA 01569 Tel 508 278 7877 800 847 6472 Fax 508 278 4555 p 1 OPHTHALMIC CRYOUNIT CR4000 OPERATIONAL INSTRUCTION MANUAL...

Page 2: ...er Hose CR4160 6 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 7 3 8 CR4000 Instructional Video CR40x 8 3 9 Ophthalmic Probes 8 4 0 Initial Set Up Procedure 8 5 0 Foot Switch Operati...

Page 3: ...t which can obscure the surgeon s view The MIRA CR4000 has a rapid freeze and defrost cycle available providing greater control of application time Temperature settings for both CO2 and N2O allow for...

Page 4: ...or changed When a cylinder is heated above room temperature it may exceed maximum pressure 900psi red zone In this situation allow the cylinder to equilibrate to room temperature Approx 70 F before us...

Page 5: ...o Carriers that come with factory installed sintered metal filters With a Micro fiber or sintered metal Filter assembly attached For use with basic CR4000 systems IMPORTANT Micro fiber or sintered met...

Page 6: ...ures as described in section 6 0 Sintered metal Filter element and O Ring replacement procedure CR5350 Proceed as in items 1 2 and 3 above 4 Remove the filter element 5 Inspect the sealing ring and re...

Page 7: ...r N2O CR4290 T Yoke assembly for E 4in Cylinder CO2 CR4390 T Yoke assembly with Micro filter for E 4in Cylinder CO2 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 The CR4150 is specia...

Page 8: ...ues The handles present an unobstructed view of the probe shaft and remain comfortable to the hand during use TABLE II Probe Model Numbers Model No Description Tip Diameter Ice Ball Size CR4010 Curved...

Page 9: ...er hose to the rear panel fitting labeled System Exhaust Connect the other end to an aspiration system or equivalent Note Use of the scavenger hose will result in quieter operation of the CR4000 Check...

Page 10: ...he 5 C temperature setting for Carbon Dioxide CO2 6 3 4 Depress the foot pedal for 10 seconds and then release 6 3 5 Wait 60 seconds while the exhaust gas is escaping from the rear scavenger port IMPO...

Page 11: ...for repair WARNING Do not alter the shape of the probe tip The shape of the probe tip is vital to its proper function Any change to the probe tip will void the manufacturer s warranty 8 2 Cleaning The...

Page 12: ...s completely non electric there are no electrical shock hazards The micro fiber filter Sintered metal filter assemblies prevent foreign particles or oils from entering the CR4000 console and probes Re...

Page 13: ...efore patient treatment Turn the gas cylinder OFF and depressurize the CR4000 system before changing the Micro filter Sintered metal filter or O ring When the probe is not in use the protective cap sh...

Page 14: ...er Place all probes in plastic bags Enter a brief note that includes the following Return Goods Authorization RGA Your full address for return of shipment Please include specific department Descriptio...

Page 15: ...ncurred by the buyer with the purchase or operation of the equipment or components thereof Repair to damaged parts will be made at a minimal charge The buyer shall pay all shipping charges on the equi...

Reviews: