MIR 911003 User Manual Download Page 12

MiniSpir Light

 

User manual

 

 

 

 

MINISPIR LIGHT User manual 

Rev 1.7   

page 12 di 13 

 

 

BLOW OUT ALL AIR IN LUNGS 

 

Between two tests, 

MiniSpir Light 

evaluates the repeatability of the following parameters: 

 
If FVC is > 1.0 L then: 
FEV1 

repeatable when the difference between the two largest FEV1 is ≤ 150 mL; 

FVC 

repeatable when the difference between the two largest F

VC is ≤ 150 mL; 

 
if FVC is ≤ 1.0 L then: 
FEV1 

repeatable when the difference between the two largest FEV1 is ≤ 100 mL; 

FVC 

repeatable when the difference between the two largest FVC is ≤ 100 mL; 

 
3.  DATA TRANSMISSION 

 

WARNING

 

 

Read  the  instructions  carefully  before  starting  the  transmission  of  data  taking  due  care  in  ensuring  that  all  the 
information has been properly understood. 

 

3.1 

Transmission with USB cable 

 
All data in the 

MiniSpir Light 

is transferred

 

through a USB cable connection. Refer to Paragraph 2.1 of this Manual to connect 

the device to a PC. The data measured by 

MiniSpir Light 

during a spirometry test are sent to the PC in digital form and managed 

by the winspiroLIGHT software. 
 

WARNING

 

 

Do  not  disconnect  MiniSpir  Light  from  the  PC  during  a  test.  Before  to  disconnect  MiniSpir  Light  from  the  PC  close 
winspiroLIGHT software. To disconnect MiniSpir Light remove the USB cable from the PC connector. For more details 
read the winspiroLIGHT user manual. 

 

3.2 

Upgrade Internal software 

 
MiniSpir  Light

  software can  be  upgraded  when  connected  to  a  PC via  USB

.

  Upgrades  can  be  downloaded by  registering  on 

www.spirometry.com. For further information on upgrading software see the winspiroLIGHT software manual. 
 
 

4.  MAINTENANCE 

 

WARNING

 

 

No part can be subjected to maintenance during use. 

 
 

MiniSpir Light

 is an instrument that requires very limited maintenance. The only operations to perform periodically is the  

 

  Changing the single-patient disposable turbine at each test 

 
The  maintenance  ope

rations  set  forth  in  the  User’s  Manual  must  be  carried  out  carefully.  Failing  to  observe  the  instructions 

contained in the manual may cause errors in measurement or in the interpretation of measured values. 
Modifications, adjustments, repairs, and reconfiguration must be carried out by the manufacturer or authorised persons. 
In case problems arise do not attempt to personally repair the unit. 
The setting of configuration parameters must be carried out by qualified personnel. In any case the risks  pertaining to incorrect 
settings do not constitute a hazard for the patient.

 

 

5.  PROBLEM SOLVING 

 

PROBLEM 

MESSAGE 

POSSIBLE CAUSES 

REMEDY 

MiniSpir  Light  does  not  connect  with  the 
PC

 

The USB cable is not 
correctly connected 

Check the correct connection of the 
USB cable side PC and side device. 

The driver doesn’t work 
correctly 

Check the presence of the device in 
the list of USB devices connected. 
Try to remove and connect the 
device. 

Spirometry data at the end of the  test are 
not acceptable 

The turbine don’t rotate 
correctly 

use a new turbine 

 

The test is performed in 
a wrong way 

Repeat the test following the 
indications on the screen 

 
 

Summary of Contents for 911003

Page 1: ...MiniSpir Light User manual MINISPIR LIGHT User manual Rev 1 7 page 1 di 13 User Manual Rev 1 7 Issued on 09 10 2020 Approved on 09 10 2020...

Page 2: ...recognised by the PC The device can then be used with the winspiroLIGHT software Keep the original packaging In the event that your device requires attention then always use the original packaging to...

Page 3: ...1 2 4 Device 6 1 2 5 Warnings for use in electromagnetic environments 6 1 3 Unforeseen errors 6 1 4 Labels and symbols 7 1 4 1 Identification label 7 Warning symbol for the WEEE 7 1 4 2 ESD Electrosta...

Page 4: ...quires installation by qualified personnel 1 1 5 Subject effect on the use of the device A spirometry test should only be carried out when the subject is at rest and in good health and thus in a suita...

Page 5: ...liquids Do not allow dust or foreign bodies to enter the turbine sensor in order to avoid incorrect functioning and possible damage The presence of any impurities such as hair sputum threads etc withi...

Page 6: ...ngs for use in electromagnetic environments Due to the increasing number of electronic devices computers cordless phones cell phones etc medical devices may be subject to electromagnetic interference...

Page 7: ...to return the device without charge to the dealer or distributor when a new equivalent device is purchased Due to the materials used in the manufacturing of the device disposing it as a normal waste...

Page 8: ...from a conductive material of equivalent size Some well known insulators are common plastics and glass An insulator will hold the charge and cannot be grounded and conduct the charge away Both conduct...

Page 9: ...online manual of the WinspiroLIGHT Software MiniSpir Light is able to make FVC VC IVC tests calculates an index of test acceptability quality control plus reproducibility of the spirometry tests carri...

Page 10: ...Radio Frequency Immunity 3V m 80 2700 MHz MIR will make available on request circuit diagrams component part lists descriptions calibration instructions or other information that will assist service...

Page 11: ...rately severe severe very severe Through an analysis applied to some of the indices and parameters calculated in the FVC test MiniSpir Light produces a variety of quality control comments useful to un...

Page 12: ...nnected to a PC via USB Upgrades can be downloaded by registering on www spirometry com For further information on upgrading software see the winspiroLIGHT software manual 4 MAINTENANCE WARNING No par...

Page 13: ...alteration adjustment modification or repair has been carried out by personnel not authorised by MIR If the fault is caused by lack of or incorrect routine maintenance of the machine If the machine ha...