background image

 

 

USER MANUAL 

 

ULTRA-LOW POWER INFRARED GAS SENSOR MIPEX-04-X-XX-3.1 

 

Mipex-04 User Manual.docx 

Revision 06 

March 25, 2020 

Page 1 of 41 

THE SOLE PROPERTY OF SIA MIPEX. ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED. 

 

 

ULTRA-LOW POWER 

 INFRARED GAS SENSOR 

 MIPEX-04-X-XX-3.1 

USER MANUAL 

ESAT.100100.00.06 UM 

  

 

Summary of Contents for MIPEX-04-1-00-3.1

Page 1: ...3 1 Mipex 04 User Manual docx Revision 06 March 25 2020 Page 1 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED ULTRA LOW POWER INFRARED...

Page 2: ...to 16 25 0 05 0 25 Updated response description for commands X and DATAE2 see Appendix C 2 1 Document ID number has been changed from ESAT 100000 00 XX UM to ESAT 100100 00 XX UM 04 April 8 2019 Upda...

Page 3: ...MIPEX 04 X XX 3 1 Mipex 04 User Manual docx Revision 06 March 25 2020 Page 3 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Rev Date C...

Page 4: ...e and transporation 15 8 Warranty 16 9 Contacts 17 Appendix A Sensor types and characteristics 18 Appendix B Connection diagrams 24 Appendix C Sensor communication protocol 26 Appendix C 1 General inf...

Page 5: ...ss I Division 1 Group A B C D according to UL Std 913 CAN CSA Std C22 2 No 157 92 Any command stated in this UM as X where X stands for a command text consisted of any number of characters must be rea...

Page 6: ...wavelength range of infrared radiation absorbed by gases while the other one detects radiation only in the wavelength range of 3 5 3 7 m Gas flowing through the cell absorbs the radiation of the oper...

Page 7: ...han years 7 IP rating 20 without dust filter 54 if a dust filter provided by SIA MIPEX is applied see Appendix E Table 3 Measurement specifications Typical measurement range vol 0 2 5 0 5 0 100 Basic...

Page 8: ...ication when the logo is present on the sensor The details of IECEx ATEX certification are specified in Appendix B see Fig 8 Sensor complies with CU TR 012 2011 Ex ia I U Ex ia IIC U according to CU T...

Page 9: ...required electrical clearances and creepage paths in accordance with EN 60079 11 Combined intrinsically safe sensor circuits parameters IECEx ATEX Ui 5 5 V Ii 200 mA Pi 0 13 W Ci 26 F Li 0 CAN CSA Vma...

Page 10: ...mmand more often than one time per 2 second 0 5 Hz may reduce sensor accuracy and increase its current consumption Correct measurement is provided when ambient temperature changes not faster than 0 6...

Page 11: ...ve it at 10 35 C for at least an hour Remove the packaging Make sure there are no mechanical injuries on sensor surfaces 6 2 Mounting Sensor should be mounted with its contacts soldered to mating PCB...

Page 12: ...for continuous 24 7 operation End user equipment UART transceiver should meet the requirements of standards IEC 60079 0 IEC 60079 11 Sensor UART transceiver communication properties are High logic le...

Page 13: ...N OF SIA MIPEX IS PROHIBITED Fig 2 Typical voltage drop within a single operating period 3 3 V Input R 100 Ohm Fig 3 Typical voltage drop within a single operating period and while data transferring 3...

Page 14: ...isplayed as 32767 hex 7FFF Sensor updates information about concentration every 1 32 0 04 seconds depends on ambient temperature and whether UART is active or not Correct measurement is provided when...

Page 15: ...ATION Transportation can be done by all means of transportation in covered vehicles as well as in a heated pressurized plane compartments in accordance with the rules of cargoes transportation actual...

Page 16: ...s caused by exposure to extremely high or low temperatures corrosion oxidation ingression of any foreign objects materials liquids insects if sensor was damaged by using connection circuit which not c...

Page 17: ...h 25 2020 Page 17 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED 9 CONTACTS MIPEX TECHNOLOGY SIA MIPEX 2B Valkas iela Daugavpils Latvia...

Page 18: ...UCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED APPENDIX A SENSOR TYPES AND CHARACTERISTICS See Table 5 for details See Table 6 for details MIPEX 04 B RX 3 1 Interface options 1 UART...

Page 19: ...1 00 3 1 CH4 CH4 0 2 5 10 40 00 MIPEX 04 1 10 3 1 0 5 10 MIPEX 04 1 20 3 1 0 100 20 MIPEX 04 1 01 3 1 0 2 5 40 60 01 MIPEX 04 1 11 3 1 0 5 11 MIPEX 04 1 21 3 1 0 100 21 MIPEX 04 1 02 3 1 0 2 5 20 50 0...

Page 20: ...at 20 RH 50 RH 90 RH 0 2 vol or 10 of indication whichever is greater within the range of 10 20 C and 25 40 C 0 4 vol or 20 of indication whichever is greater within the range of 40 10 C and 40 60 C C...

Page 21: ...ical sensitivity of MIPEX 04 1 XX 3 1 target and calibration gas is CH4 to other hydrocarbons Fig 5 Typical sensitivity of MIPEX 04 2 XX 3 1 target and calibration gas is C3H8 to other hydrocarbons 0...

Page 22: ...evision 06 March 25 2020 Page 22 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Table 7 Sensor pinout Pin Purpose 8 2 8 5 0 VDC 1 3 7 G...

Page 23: ...XX 3 1 Mipex 04 User Manual docx Revision 06 March 25 2020 Page 23 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Fig 7 Recommended ma...

Page 24: ...nsic safety control drawing MIPEX 04 X XX 3 1 Combined input parameters IECEx ATEX Ui 5 5 V Ii 200 mA Pi 0 13 W Ci 26 F Li 0 CAN CSA Vmax 5 5 V Imax 200 mA Pmax 0 13 W Ci 26 F Li 0 NET NET 8 Power 4 T...

Page 25: ...04 X XX 3 1 Mipex 04 User Manual docx Revision 06 March 25 2020 Page 25 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Fig 10 Example...

Page 26: ...parity checking Electrical parameters of UART transceiver are pointed in section 6 3 General command format any command stated in this UM as X where X stands for a command text consisted of any numbe...

Page 27: ...t syntax USER mode OEM mode Response description Command description yes yes 2 bytes in unsigned integer format see Table 8 Returns sensor readings of scaled concentration C1 not periodically single t...

Page 28: ...ata C1H7 C1L hex 0x40 C1H C1L Table 9 Structure of response to DATA command Command DATA hex 44 41 54 41 0D Byte number 1 5 6 Byte data C1 hex 0x0D Table 10 Structure of response to DATAE2 command Com...

Page 29: ...faster than by 0 6 C min Do not perform span calibration or zeroing 5 22 Dynamic temperature mode tempera ture changes faster than by 2 C min Do not perform span calibration or zeroing 6 40 Exceeding...

Page 30: ...for switching com mands level from USER to OEM XXXX stands for old password and YYYY stands for new one New password should contain only digits 0 9 in ASCII format UART yes yes USER or OEM Returns act...

Page 31: ...value incre mented to 10000 for example to write the scale coefficient of 0 009 send CALB2 00090 This com mand can be used to recalibrate sen sor if scale coefficient is known The scale coefficients...

Page 32: ...sets sensor to the factory calibration settings SETC XXXXX yes yes SETC XXXXX OK or SETC XXXXX FAULT Intended for determining the current ambient temperature in Celsius degrees XXXXX stands for degree...

Page 33: ...OK Writes a number YYYYY 00000 99999 to cell XX 00 09 of sensor memory user section Table 12 Status word displaying in concentration field while INDSIG mode is active Concentration value in ASCII for...

Page 34: ...owing gas less than 99 kPa and more than 103 kPa temperature changes faster than by 0 6 C min dust ingress if dust filter is not used data request rate more than 1 Hz 4 Recommended gas feeding rate wh...

Page 35: ...n more than 0 2 vol of target gas 4 3 if these conditions are not met sensor responds CALB AAAA FAULT 5 Stop feeding CGM 2 into sensor atmosphere 6 Feed CGM 3 into sensor atmosphere and check sensor r...

Page 36: ...egister or in the Standard designation 1 N2 0 ISO 2435 73 2 C3H8 N2 0 85 50 0 05 0 015 5328 90 3 C3H8 N2 1 6 94 0 1 0 05 EM 06 01 648 Appendix D 2 Temperature dependence of zero adjustment It is allow...

Page 37: ...MIPEX 04 X XX 3 1 Mipex 04 User Manual docx Revision 06 March 25 2020 Page 37 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Fig 11 Sc...

Page 38: ...ane Vladipor MFFC TC 6 55 221 1413 2004 On the filter back side there is an adhesive layer 3M Double Linered Laminating Adhesive 7952 in hatched areas in Fig 12 Fig 12 Dust filter outline all dimensio...

Page 39: ...e with periodic responses to command F which is intended exclusively for detailed diagnostics It returns comprehensive information on sensor readings and its status Sensor power consumption stated in...

Page 40: ...lation symbol 20 24 Uref Reference signal in ADC counts 25 Hex 0x09 Tabulation symbol 26 30 Stz0 The ratio St considering ZERO and ZERO2 coefficients 31 Hex 0x09 Tabulation symbol 32 36 Stz The ratio...

Page 41: ...s allowed A possibility of exposure to moisture Dry sensor After drying if this status lasts longer than for 20 minutes at stable conditions replace sensor 31 8 The ratio Stz0 exceeds allowable upper...

Reviews: