8.  There are two types of warning indicators on the 

HF100H

:  ERROR and WARM-UP.  If 

either one of this is lighted, x-ray cannot be generated.  

 
a. WARM-UP indicator: The internal circuits have a warm-up time of approximately 10 

seconds.  The WARM-UP light will turn on after each exposure.  

 

b. ERROR indicator: This indicates a malfunction of the 

HF100H

.  If the ERROR light is 

on, release the exposure switch.  An unusual situation has 
occurred on the inverter circuit.  Turn the main POWER switch 
OFF and wait three (3) minutes before turning the power ON 
again. If the ERROR indicator remains lighted, please consult a 
qualified service technician or contact MinXray.   

 

9.  The exposure switch has a 2-stage button.   
 

The first stage pre-heats the filament for the x-ray tube.  When this stage alone is 

depressed and held, the STAND BY indicator will light for approximately 1.5 
seconds while the filament heats.  When the STAND BY indicator turns off, the 
filament is heated and the 

HF100H

 is ready to generate x-rays.   

 

The second stage initiates the x-ray exposure for the time set in the SEC. display on 

the control.  When the second stage is fully depressed, x-ray will be emitted, 
the X-RAY indicator will light, and an audible signal will be heard.  PRESS AND 
HOLD THE FIRST AND SECOND STAGE BUTTONS TOGETHER UNTIL THE 
EXPOSURE HAS TERMINATED.  This is a “dead man” exposure switch—x-ray 
will terminate immediately as a safety feature when the buttons are released.    

 

It is possible to depress the two stage buttons simultaneously.  When this is done, 
there is approximately a 1.5 second delay before the timed x-ray exposure begins.  

 
 

10.  After each exposure, release the exposure buttons.  The 

HF100H

 is now ready for the 

next exposure.  

 
 

11. When you are finished using the 

HF100H

, turn off the main POWER switch, disconnect 

the AC power cord, and put the stand with the 

HF100H

 attached into its correct 

configuration for transport.  

 

 
 
 
 
 
 
 
 
 
 
 

 

 

  

13

Summary of Contents for HF100H

Page 1: ...on 2 0 CAUTION Federal law restricts this device to sale and use by or on the order of a licensed physician MinXray Inc 3611 Commercial Avenue Northbrook Illinois 60062 1822 U S A Tel 1 847 564 0323 T...

Page 2: ...tatements Figures and Symbols 6 Installation Instructions Unpacking 7 Assembly Instructions 7 Pre Operational Required Testing 8 Installation Check List 11 Operation Instructions Operation 12 Safety P...

Page 3: ...tablished maximum permissible doses safety recommendations and procedures derived from the following sources A National Council on Radiation Protection Report No 33 Medical X Ray and Gamma Ray Protect...

Page 4: ...perly serviced or maintained in accordance with this manual D Equipment which has been modified or tampered with in any way E Equipment that has been connected to input voltage other than the voltage...

Page 5: ...er Cord 1 pc 3 Hand held Exposure Switch and Cord 1 pc 4 Instruction Manual X ray unit 1 pc 5 Instruction Manual Collimator 1 pc OPTIONAL COMPONENTS Mobile Stand Model XGS Series Carrying Case Model H...

Page 6: ...asure Exposure switch and cord Connector for exposure switch Fuse box Power cord Knob to adjust x ray field size horizontal Connector for power cord Collimator light on button Knob to adjust x ray fie...

Page 7: ...7 Exposure Time Adjustment Button 2 mA Indicator 8 X ray Indicator 3 Exposure Time Indicator 9 Stand By Indicator 4 kV Adjustment Button 10 Error Indicator 5 kV Adjustment Button 11 Warm Up Indicator...

Page 8: ...TOR UNLESS SAFE EXPOSURE FACTORS OPERATING INSTRUCTIONS ARE OBSERVED DANGER HIGH VOLTAGE NO USER SERVICEABLE COMPONENTS INSIDE REFER SERVICE TO QUALIFIED SERVICE TECHNICIAN FOR CONTINUED PROTECTION AG...

Page 9: ...lly calling it to attention by phone where possible and filing a written report Save all packing material for inspection by the carrier If this equipment arrives with any damage it is your responsibil...

Page 10: ...exposure and read the measured value of exposure time on the NERO 5 Measure value of exposure time at following settings with same procedure 0 08 0 1 0 12 0 2 and 1 00 sec 1 2 Instruments NERO Model...

Page 11: ...Lo 70 kV 80 kV Wheel range 50 85 kV Sensitivity Lo 90 kV 100 kV Wheel range 70 120 kV Sensitivity Lo 2 2 Instruments NERO Model 6000M x ray beam analyzer Manufactured by Victoreen Inc or equivalent 2...

Page 12: ...value of ch 2 at 50 kV 60 kV 70 kV 80 kV 90 kV and 100 kV at 0 2 sec with same procedure 3 2 Instruments Digital Storage Analog Oscilloscope 2212 Manufactured by Tektronix Holland N V or equivalent 3...

Page 13: ...V 10 70 kV 10 80 kV 10 90 kV 10 100 kV 10 kV kV kV kV kV kV kV GO NO GO Tube Current 20 mA 15 mA GO NO GO Exposure Time 0 08 sec 10 1 msec 0 1 sec 10 1 msec 0 12 sec 10 1 msec 0 2 sec 10 1 msec 1 0 se...

Page 14: ...time for the view to be taken and the imaging medium used Press the arrows below the sec display to choose the exposure time 5 Use the retractable tape measure on the collimator to position the HF100H...

Page 15: ...ely 1 5 seconds while the filament heats When the STAND BY indicator turns off the filament is heated and the HF100H is ready to generate x rays The second stage initiates the x ray exposure for the t...

Page 16: ...his equipment away from likely sources of RF radiation Maintenance Clean the HF100H as needed with a damp sponge or cloth and mild soap The following must be checked before each use Tubehead 1 Make su...

Page 17: ...ed every six 6 months 1 Check the collimator light field to x ray field alignment 2 Check the kV output refer to page 9 Peak Tube Potential 3 Check the mA output refer to page 10 Tube Current 4 Check...

Page 18: ...or rating 20 mA at 100 kV 15 20 mA at 40 kV 15 Exposure timer 0 08 4 00 sec 10 1 msec Duty cycle 1 30 1 sec On 30 sec Off Preheat time Approximately 2 sec Warm up time Approximately 10 seconds after e...

Page 19: ...DIMENSIONS Specifications subject to change without notice Dimensions subject to change without notice 17...

Page 20: ...MinXray HF100H Schematic Diagram 18...

Page 21: ...19...

Page 22: ...20...

Page 23: ...HF100H TUBE HOUSING COOLING CURVE 0 100 200 300 400 500 600 0 60 120 180 240 300 360 420 480 540 600 660 720 780 TIME MINUTES HEAT UNITS STORED IN HOUSING kJ 21...

Reviews: