5. Calibration procedure
Warning
●
Before using the new instrument or after exchanging the tie bar gage, be sure to
make calibration. If calibration is not made, the correct measured results may not
be obtained, or it may cause malfunction to the instrument and it may damage the
peripheral equipment. Moreover, even if the calibration has made already, there
may occur the similar case as above when the result is not correct. So make precise
calibration again.
Warning
●
When you connect the tester and digital volt meter, etc., please connect them
correctly before turning on the power supply.
5
−
1. Preparation
1
According to the “4. Connection method”, connect this unit and a tie bar gage correctly, and
turn on the power.
2
In case of CSA
−
591
−
99 (with automatically zero function), begin the adjustment after set the
condition to turn off that function.
VR
1
A/
Z
A/
Z
O
F
F
LE
D
1
OFF ON
PCB for auto
−
zero
The front panel was removed.
PC
board
for
automatically
zero
O
N
Switch
for
automatically
zero
O
FF
Switch
for
automatically
zero