5
4.
Connecting method
4
−
1. Layout of the terminal boards
The layout of each signal of the terminal block is shown in the following table.
:
・
Terminal block
Terminal
Description
Application
1
A
Input 1 connecting terminal (RED)
2
D
Output 1
−
1 connecting terminal
(BLU)
Tie bar gage is connected.
3
D’
Output 1
−
2 connecting terminal
(WHT)
g g
4
GND
Shield (Connected internally with
terminal No.12
)
Shield of gage is connected.
D class single grounding.
5
VOUT
+
Voltage output terminal (
+
)
Voltage output
6
VOUT
−
Voltage output terminal (
−
)
Voltage output.
7
DC24 V
+
Power supply input terminal (
+
)
DC24 V is connected
8
DC24 V
−
Power supply input terminal (
−
)
DC24 V is connected.
9
C
Input 1 connecting terminal (RED)
10
B
Output 1
−
1 connecting terminal
(BLU)
Tie bar gage is connected.
11
B’
Output 1
−
2 connecting terminal
(WHT)
g g
12
GND
Shield (Connected internally with
terminal No.4)
Shield of gage is connected.
D class single grounding.
13
BUSY
Output terminal for BUSY signal
Open collector output with approx.400 ms
as an option
14
A/Z
Connecting terminal for external
automatically zero
external automatically zero
as an option
15
COM
Common terminal for BUSY/A/Z
●
Terminal No. 4 for “GND” and terminal No.12 for “GND” are connected internally.
Take the D class single ground for No.4 or No.12.
Summary of Contents for CSA-591