4 Installation and connection information
4.1 General information
-
The safety instructions in Chapter
must be followed!
-
Only use the metal EX EMC cable glands provided by the manufacturer.
-
Install the junction box so that the cable glands are on the bottom.
-
Do not open the junction box when connected to the voltage supply.
Note:
At ambient temperatures >45°C, cables that are suitable for at least 85°C must be used.
4.2 Terminals and jumpers
Terminal contacts
LC1–8
+ supply voltage (load cells)
+ Red
Terminal contact
PS
+ supply voltage (supply cable PR 6124/xxP for
the PR 6024/62S power supply)
Terminal contacts
LC1–8
- supply voltage (load cells)
-
Blue
Terminal contact
PS
- supply voltage (supply cable PR 6124/xxP for
the PR 6024/62S power supply)
Terminal contacts
LC1–8
RS-485 TxB (data transmission)
B green
Terminal contact
IND
RS-485 TxB (data cable PR 6124/xxD for the device)
Terminal contacts
LC1–8
RS-485 TxA (data transmission)
A gray
Terminal contact
IND
RS-485 TxA (data cable PR 6124/xxD for the device)
4 Installation and connection information
Cable junction box for digital load cells PR 6024/68S
Minebea Intec
EN-11